VT-R-01, VT-R-03 heating constant temperature baths 

Operating manual 

 

 

The  operating  principle  of  the  baths  is 

based  on  supporting a preset  constant  tempera-

ture of flowing thermal fluid in the bath tank and 

providing  a  good  temperature  uniformity  of  the 

operating bath. 

Pump 5  pours  thermal  fluid  from  the  top 

part of the bath into chamber 1 and, then, to the 

bottom of the bath by means of output pipe 2. It 

provides  circulation  around  Reid-vapor  pressure 

vessels 4. To drain thermal fluid, use valve 5. 

Maintaining  of  the  preset  temperature  by 

means of heating is provided by immersion circu-

lator 1. 
 

The cooling of the thermal fluid is carried out by means of heat exchange with environment or 

cooling liquid, passed through built-in cooling coil 3. 

Environmental Conditions 

Indoor use only. 
Ambient temperature: +10...+35 

°С

Air humidity: max. relative humidity 80 % for temperatures up to +31 

°С

Max. mains fluctuation of ±10

 % are permissible. 

Safety Recommendations 

Avoid strikes to the housing, vibrations, damage to the operating element panel (keypad, dis-

play), and contamination. 

Do not store the instrument in aggressive atmosphere. 
Protect the instrument from contamination. 
Only  qualified  personnel  are  authorized  to  perform  configuration,  installation,  maintenance 

and repairs of the circulator. 

Routine operation can also be carried out by untrained personnel who should however be in-

structed by trained personnel. 

 

CAUTION: The instrument is not for use in explosive atmosphere. 

 

 

 

 
 
 

Содержание VT-R-01

Страница 1: ...HEATING CONSTANT TEMPERATURE BATHS VT R 01 VT R 03 Operating manual ...

Страница 2: ...ure baths Operating manual 2 TABLE OF CONTENTS Introduction 3 Intended use 3 Appearance and parts names 3 Environmental Conditions 4 Safety Recommendations 4 Using The Heating constant temperature Baths 5 Preparation 5 GENERAL SPECIFICATIONS 6 ...

Страница 3: ...standard test method for vapor pressure of petroleum products Reid Method according to ASTM D323 and ASTM D1267 Appearance and parts names Heating constant temperature baths consist of M03M heating immersion circulator 1 and a bath tank 2 The bath tank includes built in cooling coil 4 and drain valve 5 When running tests install Reid vapor pressure vessels with pressure gauges 3 Adapter 6 is inten...

Страница 4: ...rried out by means of heat exchange with environment or cooling liquid passed through built in cooling coil 3 Environmental Conditions Indoor use only Ambient temperature 10 35 С Air humidity max relative humidity 80 for temperatures up to 31 С Max mains fluctuation of 10 are permissible Safety Recommendations Avoid strikes to the housing vibrations damage to the operating element panel keypad dis...

Страница 5: ... to the power source Attach a pressure gauge to a Reid vapor pressure vessel Insert the vessel into adapter 6 and slide into the slot Fill up the bath tank with thermal fluid through aperture under bath cover 5 When sub merging test vessels make sure that thermal fluid level will not go over the edges of the tank Put a hose on valve socket to drain the excess of ther mal fluid When operating instr...

Страница 6: ...ability 0 1 С Temperature uniformity 0 1 С Digital setting accuracy 0 5 С Digital setting repeatability 0 05 С Heating capacity 2000 W Bath volume VT R 01 VT R 03 30 Liters 40 Liters Dimensions W D H VT R 01 VT R 03 335 280 780 mm 545 250 780 mm Bath opening VT R 01 VT R 03 120 210 mm 290 160 mm Bath depth 500 mm Weight VT R 01 VT R 03 22 kg 32 kg Power supply 230 V 50 Hz 2 2 kW Warranty 2 years ...

Отзывы: