TensCare TouchTENS XL-T Скачать руководство пользователя страница 14

23

24

Recommended separation distances between portable and 
mobile RF communications equipment and the Touch TENS

The model itouch is intended for use in an electromagnetic environment in which 
radiated RF disturbances are controlled. The customer or the user of the model itouch 
can help prevent electromagnetic interference by maintaining a minimum distance 
between portable and mobile RF communications equipment (transmitters) and the 
model itouch as recommended below, according to the maximum output power of the 
communications equipment.

Rated maximum 

Separation distance according to frequency of transmitter M

output power of
transmitter W
 

 

  150kHz to 80MHz 

80 MHz to 800 MHz  800MHz to 2.5GHz

 

 

 

d=1.2√ P 

d=1.2√ P 

d=2.3√ P

 

0.01 

 

0.12  

0.12 

 0.23

 

0.1  

 

0.38  

0.38  

0.73

 

1  

 

1.2  

1.2 

 2.3

 

10  

 

3.8  

3.8  

7.3

 

100  

 

12  

12 

 23 

For transmitters rated at a maximum output power not listed above, the recommended separation 
distance d in meters (m) can be estimated using the equation applicable to the frequency of the 
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according 
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected 
by absorption and reflection from structures, objects and people.

EC Declaration of Conformity

TensCare  Ltd  hereby  declare  that  an  examination 
of  the  production  quality  assurance  system  has 
been  carried  out  following  the  requirements 
of  the  UK  national  legislation  according  to  
Annex V of the Directive 93/42/EEC on medical devices. 
We certify that the production quality system conforms 
with  the  relevant  provisions  of  the  aforementioned 
legislation, and the result entitles the organization to use 
the CE 0473 marking on this product.

Содержание TouchTENS XL-T

Страница 1: ...Model XL Y1 T INSTRUCTIONS FOR USE...

Страница 2: ...lip 4 7 The Programmes 4 8 Programme Settings 6 9 Controls and Display 7 10 Controls 8 11 Connecting Lead wires 11 12 Positioning of Electrode Pads 12 13 General Pad Advice 14 14 Cleaning 14 15 Troubl...

Страница 3: ...energy output levels for easy use A choice of eight pre set programmes is available The operation of the controls is described in detail in Sections 9 10 The Tenscare Touch TENS is a dual channel TENS...

Страница 4: ...ispose of batteries safely Do not throw batteries onto a fire Risk of explosion Battery Life Batteries should last about 48 hours at full power Unused batteries have a nominal shelf life of 3 years bu...

Страница 5: ...00 Modulated modulating continuously this gives a pleasant slow relaxing massage sensation G Frequency This changes the frequency every 3 minutes Pain Gate 2 110 200 Modulated stepping from 2 to 110 H...

Страница 6: ...re are four coloured buttons STRENGTH CONTROLS ON OFF Channel 1 Strength DOWN Channel 1 Strength UP Channel 2 Strength UP Channel 2 Strength DOWN Time Selected Programme selector Channel 1 Strength Ch...

Страница 7: ...that the unit will continue functioning until you switch it off You can set session times of 20 40 or 60 minutes by pressing the green button repeatedly The LCD shows the session duration next to the...

Страница 8: ...ee Section 16 for information on where you should not place pads 3 Elapsed Time Press Programme and Right blue together and hold down for 3 seconds The display will show U for Usage and a number from...

Страница 9: ...of the tackiness of the pad by adding a tiny drop of water on each pad and spreading around If too much water is added the pads will become too soft then it is suggested in order to try re establish...

Страница 10: ...being connected to monitoring equipment with body worn electrode pads it may interfere with the signals being monitored Strong electromagnetic fields electrosurgery microwave cookers mobile phones may...

Страница 11: ...This will invalidate the guarantee Mix old new or different types of batteries Use this device leads electrodes and accessories other than those recommended by the manufacturer Performance may vary fr...

Страница 12: ...f the problem The guarantee does not cover accessories the batteries electrode pads or lead wires Please note that the Guarantee is invalidated if incorrect batteries have been fitted or the unit has...

Страница 13: ...e model itouch including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance 3V d 1 2 P 80M Hz t...

Страница 14: ...rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the max...

Страница 15: ...TensCare Ltd 9 Blenheim Road Epsom Surrey KT19 9BE UK Tel 44 0 1372 723434 www tenscare co uk Distributed by Pub No I XLY1T UK Version 1 1 06 13 Manufactured by...

Отзывы: