malignancy  should  be  avoided  as  it  may  stimulate  growth  and  promote 
spread of cancer cells. 

 

CAUTIONS: 

Caution 

should  be  used  if  you  have  a  bleeding  disorder  as 

stimulation may increase blood flow to the stimulated region. 

Caution

  should  be  used  if  you  have  suspected  or  diagnosed 

epilepsy as electrical stimulation may affect seizure threshold. 

Caution

 should be observed when using the 

Ova+

 at the same 

time  as  being  connected  to  monitoring  equipment  with  body 

worn electrode pads. It may interfere with the signals being monitored. 

Caution

 should  be  used following recent surgical  procedures. 

Stimulation may disrupt the healing process. 

Caution.

  Simultaneous  connection  to  high  frequency  surgical 

equipment may result in burns and damage to the stimulator.  

Caution.

  Strong  electromagnetic  fields  (electro-surgery/ 

microwave  cookers/  mobile  phones)  may  affect  the  correct 

operation of this unit. If it appears to behave unusually, move it away from 
these devices. 

Caution.

  Do  not  permit  use  by  children  unable  to  understand 

the  instructions  or  persons  with  cognitive  disabilities,  i.e.; 

Alzheimer’s disease or dementia. 

 

DO NOT PLACE ELECTRODE PADS: 

 

On skin  which  does not have normal sensation.  If the skin is  numb 
too  great  a  strength  may  be  used,  which  could  result  in  skin 
inflammation. 

 

On broken skin. The pads could encourage infection. 

 

On the front of the neck. This could cause the airway to close, giving 
breathing  problems.  May  cause  sudden  drop  in  blood  pressure 
(vasovagal response). 

 

Over the eyes. May affect eyesight or cause headaches. 

Содержание OVA+

Страница 1: ...INSTRUCTIONS FOR USE READ CAREFULLY BEFORE STARTING USE...

Страница 2: ...seconds Regulate the output intensity by pressing the or buttons Place two electrodes on each side of your lower tummy distance between the electrodes should be over 3cm Select the mode which works be...

Страница 3: ...PROGRAMMES 10 8 BATTERY 11 9 CONTROLS AND DISPLAY 13 10 CONTENTS OF PACK 14 11 SETTING UP 15 11 1 PREPARING THE OVA UNIT 15 11 2 USING THE OVA 15 12 PLACEMENT OF ELECTRODE PADS 16 13 APPLICATION OF E...

Страница 4: ...onment to treat symptoms of primary menstrual pain dysmenorrhoea The Ova may also help to relieve the pain of secondary menstrual pain due to conditions such as endometriosis fibroids or pelvic inflam...

Страница 5: ...cramping pain in the lower tummy during your periods In some women with secondary dysmenorrhoea the pain starts several days before the period begins and lasts all the way through the period This is u...

Страница 6: ...t period pain 5 HOW IT WORKS The Ova uses T E N S to relieve pain T E N S stands for Transcutaneous Electrical Nerve Stimulation The Ova machine delivers small electrical pulses to the body via electr...

Страница 7: ...the front of the neck can affect your heart rate Very strong stimulation across the chest may cause an extra heartbeat Do NOT use if you are or may be pregnant or in the first 6 8 weeks after childbir...

Страница 8: ...ction to high frequency surgical equipment may result in burns and damage to the stimulator Caution Strong electromagnetic fields electro surgery microwave cookers mobile phones may affect the correct...

Страница 9: ...ICE IN GOOD WORKING ORDER OBSERVE THE FOLLOWING ADDITIONAL CAUTIONS Caution Do not immerse your device in water or place it close to excessive heat such as a fireplace or radiant heater or sources of...

Страница 10: ...pulsing This programme can have the advantage of achieving for a time ongoing residual pain relief after the Ova has been switched off Mode 3 is a fast tapping vibrating sensation This setting uses th...

Страница 11: ...ncrease the intensity to keep feeling a strong sensation xii TENS can be used as long as it gives relief If the effect wears off turn off the unit for an hour to allow your nerves to re set xiii Do no...

Страница 12: ...low these instructions could cause the lithium ion battery to leak acid become hot explode or ignite and cause injury and or damage Do NOT expose to temperatures over 60 C 140 F Do NOT put store or le...

Страница 13: ...up slowly wait after each press to feel the effect Central button Every time the unit is switched ON the programme used in the last session is automatically loaded To change it press and hold the but...

Страница 14: ...through a number from 0 to 3 and the level and status of stimulation output through a 20 segment bar graph When the battery is low a blinking symbol is displayed to warn it should be recharged 10 CONT...

Страница 15: ...de pads are in place press and hold the button on the control unit for 3 to 5 seconds to switch the control unit on 2 When switched on for the first time the Ova will automatically select the Mode 0 A...

Страница 16: ...ads are placed correctly the Ova will work very well Therefore if after thirty minutes considerable relief has not been achieved either move the pads slightly or try one of the other pad positioning a...

Страница 17: ...er end of the pads The pigtails should be uppermost 13 APPLICATION OF ELECTRODE PADS To apply the pads simply peel off the clear film liner and the pads will adhere to the skin automatically Return th...

Страница 18: ...ad Do not remove by pulling on the pad s pigtail wire If you find that a small part of the pad overlaps the top of your thigh and that this is uncomfortable then it is possible to cut off this part us...

Страница 19: ...with an alcohol free antibacterial wipe or by wiping with a damp lint free cloth and a solution of mild soap and water Wipe dry Do not immerse your TENS machine in water Do not use any other cleaning...

Страница 20: ...d you of this Directive all affected products are now being marked with a crossed out wheelie bin symbol as depicted below To comply with the Directive you can return your old electro therapy unit to...

Страница 21: ...for continuous use Environmental Specifications Operating Storage Temperature range 5 to 40 C Humidity 15 to 93 RH non condensing Temperature range 25 to 70 C Humidity Up to 93 RH non condensing TYPE...

Страница 22: ...tery Low battery display Low batteries Charge battery No sensation Incorrect connection Have you applied both electrode pads to the Y cable to ensure a complete circuit Not strong enough Increase stre...

Страница 23: ...riod of the warranty Otherwise we will quote for any repair which will be carried out on acceptance of our quotation The benefits conferred by this warranty are in addition to all other rights and rem...

Страница 24: ...clude used electrode pads Send only the unit and lead wires Please return the unit and warranty card see page 25 at your cost to TensCare Ltd PainAway House 9 Blenheim Road Longmead Business Park Epso...

Страница 25: ...TY NAME ADDRESS POSTCODE DAYTIME TELEPHONE E MAIL MODEL DATE OF PURCHASE ATTACH PROOF OF PURCHASE DO NOT SEND IN PROBE OR ELECTRODE PADS RETAILERS NAME RETAILERS ADDRESS RETAILERS POSTCODE BRIEF DESCR...

Страница 26: ...Notes...

Страница 27: ...limited https plus google com TenscareLtdEpsom https uk pinterest com TensCareLtd https tenscareltd wordpress com https www youtube com channel UCzpik9dmLlJ3j0aHOpQ 0sg EC Declaration of Conformity Te...

Страница 28: ...Distributed by Manufactured by TensCare Ltd 9 Blenheim Road Epsom Surrey KT19 9BE UK Tel 44 0 1372 723434 www tenscare co uk Pub No I OVAP UK Version 3 2 07 19...

Отзывы: