Tens i-Pulz 2 Скачать руководство пользователя страница 24

24

i-Pulz 2

25

i-Pulz 2

Pre-set overview 

Pr

e-set

Fr

equency

Pulse 

width

Extra info

Adjustable parameters

P1

Conventional 

TENS

110 Hz

80 µs

Pulse: 

 

30 – 400 µs

Fr

eq: 

 

30 – 120 Hz

P2

Low 

Fr

equency 

TENS

2 Hz

250 µs

Pulse: 

 

30 - 400 µs

Fr

eq: 

 

1 – 30 Hz

P3

Burst

2 Hz

180 µs

Pulse sequence: 8 pulses

Pulse: 

 

30 - 400 µs

Fr

eq: 

 

1 – 5 Hz

Number of pulses: 2 – 25

P4

HAN

100 Hz

150 µs

Switches every 3 seconds

Pulse: 

 

30 - 400 µs

Fr

eq: 

 

30 – 120 Hz

2 Hz

200 µs

Pulse: 

 

30 - 400 µs

Fr

eq: 

 

1 – 30 Hz

low intensity: ON/OFF

P5

Pulse width 

modulation

110 Hz

60–100 µs

2 seconds r

epeat-time

Pulse1: 

 30 - 400 µs

Pulse2: 

 30 - 400 µs

Fr

eq: 

 

1 – 120 Hz

Repeat-time: 2 – 10 sec

P6

Fr

equency 

modulation

2–100 Hz

80 µs

5 seconds r

epeat-time

Pulse: 

 

30 - 400 µs

Fr

eq1: 

 

1 – 120 Hz

Fr

eq2: 

 

1 – 120 Hz

Repeat-time: 2 – 10 sec

P7

Modulated 

stimulation 1

100 Hz

150 µs

Ramp up time:

 

1 sec

Active time: 

 

4 sec

Ramp down time: 1 sec

Rest time: 

 

3 sec

Rest intensity: 

 

ON

Pulse: 

 

30 - 400 µs

Fr

eq: 

 

1 – 120 Hz

Ramp 

up/down 

time: 

0.5 

– 

2.5 

sec

Active time: 0 – 30 sec

Rest time: 0 – 60 sec

Rest intensity: ON/OFF

Содержание i-Pulz 2

Страница 1: ...USER MANUAL i Pulz 2...

Страница 2: ...ffects 5 Environment 6 Warnings 6 Precautions 6 Maintenance 7 IP rating 8 2 The i Pulz 2 nerve stimulator 8 General 8 Controls 8 Buttons 9 The display 10 Indication LED 11 Channels left and right 11 3...

Страница 3: ...ntents of package 22 8 Technical information 22 9 Pre set overview 24 10 Errors 26 Error numbers 26 Other indications faults 27 Irritated skin 29 11 Important information regarding EMC 29 General 29 T...

Страница 4: ...under the supervision of the parents Indications Typical prescriptions for use by qualified pain treatment specialist are Neurological pain like neuralgia blunt pain and phantom pain Muscle and joint...

Страница 5: ...ally the carotid sinus Crossing the heart section From electrodes placed on the chest and the upper back through the chest using an anterior and posterior electrode positions or other transthoracic ap...

Страница 6: ...put of the i Pulz 2 Do not use electrodes with a smaller surface than 15 cm as this can cause skin irritation or even burns Irritation of the skin can be caused by lengthy use of small electrodes and...

Страница 7: ...ve detergents and solvents like ether alcohol or abrasives Never submerge the device in water or in any other liquid In case it does get submerged get the device checked by the manufacturer or distrib...

Страница 8: ...hannel on the left and right side of the device Both channels work independently from each other and do not influence each other two different forms of stimulation can therefore be used simultaneously...

Страница 9: ...r the pre set Press the Timer button once or multiple times to set up the desired time ranging from 0 to 99 minutes and starts at 0 again By holding the Timer button pressed for 2 seconds the timer wi...

Страница 10: ...k If the buttons are not pressed after setting the intensity the keypad lock will activate after 15 seconds This function prevents any accidental increase in the intensity The key icon indicates that...

Страница 11: ...ft and right The channels are the outputs where the cables for the electrodes can be connected The channel on the left side is the left channel the channel on the right side is the right channel When...

Страница 12: ...id confusion Keep to the instructions of your treating specialist when placing the electrodes Setting up the stimulator After the electrodes are placed properly you can activate the device The device...

Страница 13: ...above 5 mA a warning sign and a cable icon will be displayed The intensity of the relevant channel will be indicated by nc See Figure 1 no connection When the intensity will be set higher than 7 5 mA...

Страница 14: ...he pre set button pressed for 2 seconds Or by turning down the intensity until it reaches 0 0 mA When the intensity reaches 0 0 mA the device will switch off after a few minutes in order to preserve t...

Страница 15: ...and PCB number By tapping on the pre set button it can step through the menu items Device information Cable tester Buzzer Time Timer step size Mask pre set Default settings To change the settings of...

Страница 16: ...e left up button the sound will be turned off with the right up button the sound will be turned on By pressing on the Timer button you will be asked if this change should be saved before the device go...

Страница 17: ...pre set If all pre sets are masked P1 will remain visible during normal use By pressing on the Timer button you will be asked if this change should be saved before the device will go back to the sett...

Страница 18: ...Down button for the Right channel See Figure 3 adjust menu To customize the parameters of these basic pre sets press on the pre set button First you can adjust the parameters of the Left channel Depen...

Страница 19: ...l will be reset to the same intensity as the main intensity level Compliance menu In the compliance menu it is possible to see the usage of the device The pre set intensity and time of each use will b...

Страница 20: ...adapter into the mains power socket Place the i Pulz 2 on the centre of the wireless charge pad The indicator LED on the i Pulz 2 will light up orange the device is now charging The indicator LED on t...

Страница 21: ...kind of electrode can be selected See the packaging or leaflet of the electrode for the used materials or contact your distributor for more information Disposable electrodes Disposable electrodes are...

Страница 22: ...ions Channels Two independent adjustable intensity Pulse type Asymmetric block square waveform two phased without a direct current component Intensity 0 100 mA current source continuously interchangea...

Страница 23: ...protects against electric shocks The serial number of the device consists of 6 numbers The first 2 numbers represent the year of production and the other 4 numbers represent the device number Informat...

Страница 24: ...120 Hz 2 Hz 200 s Pulse 30 400 s Freq 1 30 Hz low intensity ON OFF P5 Pulse width modulation 110 Hz 60 100 s 2 seconds repeat time Pulse1 30 400 s Pulse2 30 400 s Freq 1 120 Hz Repeat time 2 10 sec P...

Страница 25: ...q 1 120 Hz Ramp up down time 0 2 5 sec Active time 0 30 sec 5 Hz Ramp up time 0 sec Active time 5 sec Ramp down time 0 sec Freq 1 120 Hz Ramp up down time 0 5 2 5 sec Active time 0 30 sec P10 Depth TE...

Страница 26: ...istributor 7 8 19 20 21 Error screen All functionalities disabled Internal error detected in the output signal Turn OFF device and contact distributor Do not use the i Pulz 2 again until it is checked...

Страница 27: ...rn ON indicator LED is RED The battery is too low Put the device on the charger The device does not turn ON indicator LED is blinking RED Internal error with battery Remove the i Pulz 2 to cool down t...

Страница 28: ...if the cable is connected to electrodes Verify if the electrodes are correctly pasted to the body Warning sign blinking cable icon intensity value is and channel is disabled The output is disconnected...

Страница 29: ...information regarding EMC General Electronic devices such as portable and mobile RF communication devices and medical devices can be sensitive for electromagnetic interference of similar devices This...

Страница 30: ...sk for the patient user when an error should occur in the device For further directions regarding the EMC environment in which the device should be you can find more information in tables 1 to 4 below...

Страница 31: ...contact 15 kV air 8 kV contact 15 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transi...

Страница 32: ...Common mode Conducted Susceptibility 3 V m 80 MHz to 2 7 GHz 3 V m d 1 2 P 80 MHz to 800 MHz d 2 4 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W accordi...

Страница 33: ...r radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey...

Страница 34: ...distance according to frequency of transmitter in meters 150 kHz 80 MHz d 1 2 P 80 MHz 800 MHz d 1 2 P 800 MHz 2 5 GHz d 2 4 P 0 01 0 12 0 12 0 24 0 1 0 38 0 38 0 76 1 1 2 1 2 2 4 10 3 8 3 8 7 6 100...

Страница 35: ......

Страница 36: ...Manufacturer Distributor Rehan Electronics Ltd Aughrim Road Carnew Co Wicklow Ireland www rehanelectronics com Tel 00353 53 9426742 Fax 00353 53 9426051...

Отзывы: