Important notes to avoid damage:

 - Do not allow power supply to sink current.  (Use diode when charging battery.)
 - In series mode, the voltage at the output is the sum of the voltages on the CH1 and CH2
displays
 - In parallel mode, the current at the output is the sum of the currents on the CH1 and CH2
displays
 - In parallel mode, connect load to CH2 (right hand) channel.  (When mains power is turned
on, the power supply defaults to independent mode, and CH1 to whatever voltage and cur-
rent settings are in CH1 memory recall M1, with the load turned off.)
 - OCP and OVP do not work in serial and parallel modes.
 - For sensitive circuits, disconnect load before:
         - Switching load on or off with “OFF/ON” button for that channel
         - Switching mains off with the load switched on
Do not leave a load connected to CH2 in SER (serial) mode.  If you recall a CH2 memory in
SER mode, only CH2 is updated from that memory and only CH2 turns off.  At this time, CH2
output terminals have up to -1.2 V (i.e. reverse polarity) on them, with a short circuit current
of up to -2.8 A.

Important notes to avoid frustration:

 - Automatically switches between constant voltage (CV) and constant current (CC) modes
 - There is a small +/- voltage and current while the load is switched off (50mV, 100 uA)
 - CH2 OUT LED stays on after CH2 OCP/OVP trip disconnects the load.
 - CH2 OVP/OCP buttons

also

control CH1 OVP/OCP

 - CH1 OVP/OCP buttons

only

control CH1 OVP/OCP

 - The OCP and OVP LEDs apply to whichever channel last updated them.  Pressing either
the OVP or OCP buttons updates both the OVP and OCP LEDs with current state for that
channel.

Edits are in colour, and from an end user who has no affiliation with the supplier or manufac-
turer.  No liability is accepted for any information in, or omission from, this document.  Your
unit may work differently than described here if it has a different hardware or software ver-
sion.

Use at your own risk

 - TENMA 72-10495 (Korad KA3005D-2S):  30V-5A x 2 Channels
 - TENMA 72-10500 (Korad KA3003D-2S):  30V-3A x 2 Channels
 - TENMA 72-10505 (Korad KA3003D-3S):  30V-3A x 2 Channels & 5V-3A x 1

Velleman supply the single channel version with remote interface as the Velleman LABPS3005D
(which they used to call the PS3005D)  The "D" in the Korad model number means no remote in-
terface.

Содержание KA3003D-2S

Страница 1: ...rent of up to 2 8 A Important notes to avoid frustration Automatically switches between constant voltage CV and constant current CC modes There is a small voltage and current while the load is switched off 50mV 100 uA CH2 OUT LED stays on after CH2 OCP OVP trip disconnects the load CH2 OVP OCP buttons also control CH1 OVP OCP CH1 OVP OCP buttons only control CH1 OVP OCP The OCP and OVP LEDs apply ...

Страница 2: ......

Страница 3: ......

Страница 4: ...5 pairs ...

Страница 5: ...ast pressed for Pressing ei ther the OCP or OVP button for a channel will toggle its state and up date both the OCP and OVP LEDs If you press and hold the OCP BEEP button for 2 secs to turn beep on off it inadvertently changes the state of channel 1 OCP Likewise pressing the PARA and SER but tons for 2 seconds will in advertently change the state of channel 2 OCP or OVP when entering in dependent ...

Страница 6: ...When beep is turned off some keys may still occasionally beep When beep is turned on an OCP or OVP trip will make the unit beep once If you press and hold the OCP BEEP button for 2 secs to turn beep on off it inadver tently changes the state of the OCP ...

Страница 7: ...s The voltage at the out put is the sum of the voltages on the CH1 and CH2 displays SER PAR M5 M5 OVP and OCP do not work in series and parallel modes even though their LEDs might be on When you exit serial or parallel mode the CH2 values remain repeated on the CH1 side and one memory LED is on for CH1 but the values on the CH1 display while being the currently used set values for CH1 are not actu...

Страница 8: ...mode this is the desired constant voltage For use in constant current mode this is the desired voltage limit If you enable overvoltage protection the output switches off automatically if the volt age exceeds this value Current setting For use in con stant voltage mode this is the desired current limit For use in con stant current mode this is the desired constant current If you enable overcurrent ...

Страница 9: ... OVP OCP CH1 OVP OCP buttons only control CH1 OVP OCP Pressing either the OCP or OVP button for a channel will toggle its state and update both the OCP and OVP LEDs with the OCP OVP states associated with that channel The above facts lead to some interesting but logical behavior You have to know what channel last updated those LEDs to know what channel the LEDs apply to Another example The OCP and...

Страница 10: ......

Страница 11: ... Zero Calibration connect the positive pole and Negative pole of the multimeter leads into the output terminals respectively Then watch the voltmeter and adjust the knob to make the multimeter display in the range 0V 5mV After that the zero calibration is over Press the button M1 to save the calibration value 2 C u r r e n t Z e r o C a l i b r a t i o n p r e s s t h e b u t t o n VOLTAGE CURRENT...

Отзывы: