72-9490: OPERATING MANUAL
The double display shows the first phase value of power factor PF and phase angle PG.
To test for Power factor (main display) + Phase angle (secondary display), connect the Meter as follows:
1. Press the
MENU
to select Power factor (main display) + Phase angle (secondary display) range.
2. Press the lever to open the transformer jaw, and clamp them to the corrresponding phase of tested conductor. If
user needs to mesaure any phase of the 3 phase, then clamp them to that phase’s conductor.
3. The connecting method of 3 phases 4 wires, 3 phases 3 wires or single phase 2 wires, see figure 6, 7 and 8
4. When measuring 3 phase 4 wires: (see figure 20, 21, 22)
Press
SELECT
to choose the first phase 1, see figure 20.
To avoid damage to the Meter or personal injury, do not exceed AC voltages of 600V rms and AC
Current of 1000A rms.
Warning
E. Power Factor (main display) + Phase Angle (secondary display) Measurement
Figure 20
Then press
SELECT
again to choose the second phase 2
,
see figure 21.
Figure 21