Measurement Operation(8)
F
.
Temperature Measurement
(see figure 8)
The temperature measurement ranges are -40
o
C~1000
o
C
and -40
o
F~1832
o
F.
To measure temperature, connect the Meter as follows:
1. Insert the red temperature probe into the
V
Ω
terminal and the black temperature probe into the
COM
terminal.
2. Set the rotary switch to
o
F
o
C and press
SELECT
button
to select
o
F or
o
C measurement mode.
o
C measurement
mode is default.
3. Place the temperature probe to the object being measured.
The measured value shows on the display.
red black
( figure 8)
Note
l
The Meter automatically displays the temperature
value inside the Meter when there is no temperature
probe connection.
Model 72-7222: OPERATING MANUAL
22
Содержание 72-7222