MEASUREMENT DIRECTION OF THE ROTATING MAGNETIC FIELD
It is necessary to measure the direction of the rotating magnetic field in the following
way:
•
Insert terminals L1, L2 and L3 of the test pen into the L1, L2 and L3 holes
respectively.
•
Insert the other terminal of the test pen into the alligator clip.
•
Allow the alligator clip access to the phases of the three power cables to be
measured.
•
Afterwards, the LCD will automatically display the sequence of L1, L2 and L3.
SPECIFICATION
Environment
Working temperature
0°C ~ 40°C (32°F ~ 104°F)
Storage temperature
0°C ~ 50°C (32°F ~ 122°F)
Elevation
2000m
Humidity
≤95%
Pollution protection grade
2
IP grade
IP40
Mechanical Specification
Dimensions
123 x 71 x 29mm (4.8 x 2.8 x 1.1in)
Weight
160g
Safety Specification
Electrical safety
Comply with IEC61010/EN61010 & IEC61557-7
Max. operating voltage (Ume)
700V
CAT grade
CAT III 600V
Electrical Specification
Power supply
Provided by the measured device
Nominal voltage
40VAC - 700VAC
Frequency (fn)
15Hz - 400Hz
Current Induction
1mA
Nominal test current (subject to
each phase)
1mA
Содержание 72-3515