Resistance
400.0
¡
± (1.0% + 4 digits)
4.000K
¡
± (1.5% + 2 digits)
40.00K
¡
400.0K
¡
4.000M
¡
± (2.5% + 3 digits)
40.00M
¡
± (3.5% + 5 digits)
Capacitance
40.00nF
r
(4.0% r 20 digits)
400.0nF
r
(3% r 5 digits)
4.000
F
40.00
F
100.0
F
r
(4.0% r 10 digits)
Frequency
10-10kHz
Sensitivity:100V(<50Hz);50V(50 to
400Hz;15V(401Hz to 10kHz)
r
(1.5% r 2 digits)
Temp
(type-K)
(probe accuracy
not included)
-20.0 to 760.0
q
C
r
(3%rdg+5
q
C)
-4.0 to1400.0
q
F
r
(3%rdg+9
q
F)
Clamp size
Opening 1.2" (30mm) approx
Diode Test
Test current of 0.3mA typical; Open circuit voltage 1.5V DC
typical.
Continuity Check
Threshold <150
:
; Test current < 0.5mA
Low Battery Indication
“ ” is displayed
Overrange Indication
“OL”” is displayed
Measurements Rate
2 per second, nominal
Input Impedance
10M
:
(VDC and VAC)
Display
4000 counts LCD
AC Current
50-60Hz (AAC)
AC Voltage bandwidth
50-400Hz (VAC)
Operating Temperature
41 to 104
o
F (5 to 40
o
C)
Storage Temperature
-4 to 140
o
F (-20 to 60
o
C)
Operating Humidity
Max 80% up to 87
q
F (31
q
C) decreasing linearly to 50% at
104
q
F(40
q
C)
Storage Humidity
<80%
Operating Altitude
7000ft. (2000meters) maximum.
Over voltage
Category III 600V
Battery
One 9V Battery
Auto OFF
approx. 30 minutes
Dimensions/Weight
197x70x40mm/183g
Safety
For indoor use and in accordance with Overvoltage Category
II, Pollution Degree 2. Category II includes local level,
appliance, portable equipment, etc., with transient
overvoltages less than Overvoltage Cat. III
Operation
NOTICES
: Read and understand all
warning
and
precaution
statements listed in the
safety section of this operation manual prior to using this meter. Set the function
select switch to the OFF position when the meter is not in use.
AC/DC Current Measurements
WARNING:
Ensure that the test leads are
disconnected from the meter before making current
clamp measurements.
1.
Set
the
Function
switch
to
the
400ADC,40ADC,400AAC or40AAC
range.
2. If the range of the measured is not known,
select the higher range first then move to the lower
range if necessary.
3. Press the trigger to open jaw. Fully enclose one conductor
to be measured.
The clamp meter LCD will display the reading.
AC/DC Voltage Measurements
1. Insert the black test lead into the negative
COM
terminal and the
red test lead into the positive
V
terminal.
2. Set the function switch to the
V
position.
3. Select AC or DC with the
MODE
button.
4. Connect the test leads in parallel to the circuit under test.
5.
Read the voltage measurement on the LCD display.
Resistance Measurements
1.
Insert the black test lead into the negative
COM
terminal and the
red test lead into the positive terminal.
2.
Set the function switch to the
:
•))) CAP
position.
3.
Touch the test probe tips across the circuit or component under
test. It is best to disconnect one side of the device under test so
the rest of the circuit will not interfere with the resistance reading.
4.
For Resistance tests, read the resistance on the LCD display.
Diode and Continuity Measurements
1. Insert the black test lead banana plug into the negative
COM
jack and the red test lead
banana plug into the positive diode jack.
2. Turn the rotary switch to the
:
•)))
CAP position.
3. Press the
MODE
button until “ “ appears in the display.
4. Touch the test probes to the diode under test. Forward voltage will indicate 0.4V to 0.7V.
Reverse voltage will indicate “
OL
”. Shorted devices will indicate near 0mV and an open
device will indicate “
OL
” in both polarities.
Red Black Black Red
Probe Probe Probe Probe
Forward test Reverse test
¡
¡
Operation
NOTICES
: Read and understand all
warning
and
precaution
statements listed in the
safety section of this operation manual prior to using this meter. Set the function
select switch to the OFF position when the meter is not in use.
AC/DC Current Measurements
WARNING:
Ensure that the test leads are
disconnected from the meter before making current
clamp measurements.
1.
Set
the
Function
switch
to
the
400ADC,40ADC,400AAC or40AAC
range.
2. If the range of the measured is not known,
select the higher range first then move to the lower
range if necessary.
3. Press the trigger to open jaw. Fully enclose one conductor
to be measured.
The clamp meter LCD will display the reading.
AC/DC Voltage Measurements
1. Insert the black test lead into the negative
COM
terminal and the
red test lead into the positive
V
terminal.
2. Set the function switch to the
V
position.
3. Select AC or DC with the
MODE
button.
4. Connect the test leads in parallel to the circuit under test.
5.
Read the voltage measurement on the LCD display.
Resistance Measurements
1.
Insert the black test lead into the negative
COM
terminal and the
red test lead into the positive terminal.
2.
Set the function switch to the
:
•))) CAP
position.
3.
Touch the test probe tips across the circuit or component under
test. It is best to disconnect one side of the device under test so
the rest of the circuit will not interfere with the resistance reading.
4.
For Resistance tests, read the resistance on the LCD display.
Diode and Continuity Measurements
1. Insert the black test lead banana plug into the negative
COM
jack and the red test lead
banana plug into the positive diode jack.
2. Turn the rotary switch to the
:
•)))
CAP position.
3. Press the
MODE
button until “ “ appears in the display.
4. Touch the test probes to the diode under test. Forward voltage will indicate 0.4V to 0.7V.
Reverse voltage will indicate “
OL
”. Shorted devices will indicate near 0mV and an open
device will indicate “
OL
” in both polarities.
Red Black Black Red
Probe Probe Probe Probe
Forward test Reverse test
¡
¡