G. Frequency Measurement
(see figure 9)
(Figure 9)
Warning
To avoid the risk of personal injury, do not attempt to measure frequency with voltage higher than 30V RMS.
To measure frequency, connect the Meter as follows:
1. Insert the red test lead into the
Hz
terminal and the black test lead into the
COM
terminal.
2. Set the rotary switch to
Hz
0
F
and press
SELECT
button to select
Hz
measurement mode
3. Connect the test leads across with the object being measured. The measured value shows on the display.
Note
• When making frequency measurements, the measured signal “a”, must fall within the following
voltage level:
When 10Hz to 1MHz
: 150m ≤ a ≤ 30 rms
When > 1MHz to 10MHz : 300m ≤ a ≤ 30 rms
When > 10MHz to 50MHz : 600m ≤ a ≤ 30 rms
When > 50MHz
: Unspecified
• When frequency measurement is complete, disconnect the test leads from the circuit under test, and remove from the
input terminals of the meter.