Soldering Station
User’s Manual
5. Temperature calibration
You need calibrate the temperature of tip after you replace with a new heating element or tip.
5.1 Enter into calibrating station by long pressing knob “
*
” (>3s).
5.2. You can directly adjust the value of calibration by pressing knob “▲” or “▼”.
5.3. The value of calibration is temperature measure minus the setting. (e.g. Actual value 380°C - Setting value
350°C = +30°C. Pressing knob “▲” adds 30°C; Actual value 320°C - Setting value 350°C = 30°C.
Pressing knob “▼” minus 30°C)
5.4. The calibrating temp range is +50°C to -50°C.
5.5. You can press knob “
*
” to store after you finish calibration. (Diagram 4)
6. Temp. unit exchange
In the power off condition, press and hold knob “#”, then turn on the station, the temp unit will be changed
between “°C” and “°F” and store automatically
7. False alarm
When “H-E” or “S-E”is displayed on the screen, there is some wrong in heating element or the circuit.
(Diagram 5,6). Turn off the unit and follow the instructions to replace the heating element.