background image

 

 

TEN 

 
 

 

Test Equipment Nederland B.V. (C)      Version 3.0  February 2008 

Pagina 8

 

 

CS_A CS_B

 

The corresponding checksum of the software. 

 

SERIAL 

The serial number of the smoke head.  

 

CURR.DATE 

Current date.  

  

CAL.DUE 

Calibration due date. 

 

CNT.   

Calibration counter. After each calibration this counter automatically increments. 

  

FILTER 

 

Dynamic test to check the low-pass filter-time constant. First a zero calibration is performed. After 
this the EDA 2 checks the rise time from 0 till 90% opacity. Than the EDA displays ‘

time 1.00

’. This 

value must be between  0,95 and 1,05 seconds. Normally the time is  1,00 seconds. 

 

 

During this test  the sample probe must be removed from the exhaust. Don’t influence the smoke  

 

 

meter in any way!  

 
Remark: the  ‘STATUS’ menu is only used for checking purposes.   
 
 
 

  4  Official test

 

 
 
During the smoke test just press the Enter key to start the official test. The official test is a series of 3 free 
accelerations to measure the maximum emission of a diesel vehicle. 
 
First the EDA 2 performs a zero calibration. Be sure that during a zero calibration the sample is removed from the 
exhaust tail pipe. After the zero calibration you insert the probe in exhaust of the vehicle.  
 
Then the oil temperature is measured. The EDA 2 waits until a temperature of 60  ºC is reached. As soon as the 
engine temperature reaches 60  ºC, the EDA automatically continues. If the temperature of 60  ºC is not reached or 
you just want to continue, press the Enter key. Now the INNOVA 500 will proceed with the next program part. 
 
After measuring the temperature the operator must be prepared to perform the  free accelerations 
 
Between each acceleration the operator can see a down counter of 6 seconds. During the count down the throttle of 
the vehicle is not touched. 
 
After the count down the INNOVA shows ‘THROTTLE UP’ which means that the operator has to press the pedal 
completely until the maximum engine speed is reached. Maintain this situation for maximum 2 seconds. Then 
release the pedal and let the engine speed come back to idle. The SMOKE METER light blinks. 
After 4 seconds the INNOVA 500 shows ‘RELEASE THROTTLE’. These 4 seconds are the maximum throttle up 
time for vehicle with a ‘slow’ engine (large commercial vehicles). 

For passenger cars just maintain the maximum 

engine speed for 2 seconds ! 

 
After each acceleration the EDA 2 displays the result of the test for 3 seconds. After this the INNOVA 500 starts a 
new count down for the next acceleration.  
 
After completing the official test, the EDA 2 shows the result of the test, the calculated average of the opacity of the 
3 free accelerations, at the opacity display. If the result is out of limits, more than 50%, the display blinks. 
 
 
 
 
 

Содержание EDA 2

Страница 1: ...ftware are the intellectual property of Test Equipment Nederland B V Each way of duplication of this manual or software is not allowed Test Equipment Nederland B V is not responsible for any direct or indirect damage arising from the use of the apparatus or manual The EDA 2 is tested and approved according the EMC directive 89 336 EC ...

Страница 2: ...TEN Test Equipment Nederland B V C Version 3 0 February 2008 Pagina 2 Contents 1 Introduction 3 2 Preparations 4 3 Smoke test 6 4 Official test 8 5 Messages 10 6 Maintenance 10 7 Specifications 12 ...

Страница 3: ...completely made of aluminum and uses the so called unique natural flow system that guarantees a minimum of maintenance In this user manual we describe the combination of the EDA2 smoke head and the INNOVA500 The INNOVA 500 has digital LED displays and a thermal printer to provide you all necessary information during a emission test Before using the EDA 2 it s absolutely necessary to read this user...

Страница 4: ...RPM is recorded correctly If not repeat the procedure mentioned above at a different location of the injection pipe Figure 2 Placing the piëzo clamp Remark the piëzo clamp can not be used for all types of diesel vehicles for example common rail diesels Here for you need a special RPM adapter If you use an adapter always follow the instructions given by the manufacturer Oil temperature sensor Take ...

Страница 5: ...ameter equal or larger than 70 mm Place the probe a far as possible in the exhaust Take care the gas flow in probe is correct Avoid sharps curves or bending Figure 3 Placing the sample probe The temperature of the sample probe can rise up to 200 ºC during a smoke test Always wear protection gloves while you are touching the exhaust or sample probe During a zero calibration of the smoke meter the s...

Страница 6: ...displayed If the opacity and max opacity display shows the message you need to perform a zero calibration Press the zero calibration key to activate a zero calibration Figure 5 Zero calibration of the smoke head Important We advise you to perform a zero calibration regularly Take care that during a zero calibration the sample is removed from the exhaust of the vehicle Printing If you wish to print...

Страница 7: ...during a test always use MODE Select mode A not filtered or mode B filtered Normally during a test always use mode B PROBE Select sample probe 1 or 2 Probe 1 Ø 10 mm exhaust diameter 70 mm passenger cars Probe 2 Ø 27 mm exhaust diameter 70 mm large commercial vehicles DIVISION Here you can set the division ratio for the RPM for example 1 2 3 4 etc DISPLAY Select what you want to display at the dis...

Страница 8: ...l a temperature of 60 ºC is reached As soon as the engine temperature reaches 60 ºC the EDA automatically continues If the temperature of 60 ºC is not reached or you just want to continue press the Enter key Now the INNOVA 500 will proceed with the next program part After measuring the temperature the operator must be prepared to perform the free accelerations Between each acceleration the operato...

Страница 9: ...Nederland B V C Version 3 0 February 2008 Pagina 9 Press the print key to get a full report of the official test with all recorded values Press the escape key to close the official test and return to the normal smoke test ...

Страница 10: ...A 500 remove the power lead and let the smoke head cool down until ambient temperature Open the little lit of the main switch at the rear of the EDA no 12 of figure 2 Take out the fuses and place the new fuses in the holders of the lit Use 230 V 2AT fuses Close the lit In the smoke head the fuse holder is located at the bottom of the EDA 2 no 5 of figure 3 Press and open the holder by turning it c...

Страница 11: ... of the new light source at the bottom but leaf the plastic around the bulb Carefully press the new light source with its two pins at the bottom into the holder of the support Remove the plastic cover Never touch the light source with you fingers Place the support into the cone and put the cone into the smoke chamber Tighten the mounting ring Switch on the smoke meter and perform a zero calibratio...

Страница 12: ...30 mm Material Aluminum Diameter 55 mm Light source Halogen 6V 10W 3200 ºKelvin Receiver Siemens BPW21 with V filter Sample probe 10 mm 1 meter length Optional 16 mm 2 5 meter length 27 mm 1 meter length trucks 27 mm 3 5 meter length trucks Ranges Opacity 0 99 9 0 9 99 m 1 Oil temperature 0 150 ºC Cell temperature 0 150 ºC RPM 300 9990 1 min Resolution Opacity 0 1 0 01 m 1 Oil temperature 1 ºC Cel...

Отзывы: