Temspec Air Medic TIF-06 Скачать руководство пользователя страница 7

 

- 5 - 

5.1.2

 

HEPA Filter

5.1.2.1

 

Description 

The  Air  Medic  comes  standard  with  a  6-inch  HEPA  (High 
Efficiency Particulate Air) filter. This filter is designed to stop at 
least 95% of all particles 0.3µm or larger. This is same level of 
filtration as N95 masks. 

5.1.2.2

 

Maintenance Interval 

The  unit  is  equipped  with  a  filter  change  light  that  will 
indicated  when  the  HEPA  filter  should  be  changed.  The 
standard  replacement  frequency  for  the  HEPA  filter  can  be 
found in Table 1.  

5.1.2.3

 

Replacement Instructions  

To access the HEPA filter, remove the upper and lower access 
panels. The HEPA filter is held to the bulkhead by two brackets. 
These brackets can be removed by loosening the wingnuts and 
turning the brackets so that they come free of the threaded 
rods.  The  HEPA  can  then  be  lifted  and  removed.  When 
inserting the new HEPA make sure the airflow arrow is pointing 
up,  and  the  HEPA  is  sitting  flat  on  the  bulkhead.  Put  the 
brackets back in place on the threaded rods and then tighten 
the wingnuts. 

 

 

5.1.3

 

Activated Carbon Filter (Optional Accessory)

5.1.3.1

 

Description 

An optional accessory for the Air Medic is two 1-inch Activated 
Carbon Filters. These filters make use of Activated Carbon to 
adsorb airborne particles to help with allergens and odours.  

5.1.3.2

 

Installation Instructions 

The Activated Carbon filters are installed on the intakes behind 
the  Prefilters.  To  install  the  new  filter,  first  ensure  that  the 
airflow arrow on the filter is pointing towards the middle of the 
unit. Slide the filter in lower into the upper and lower guides, 
ensure it is pushed all the way to the bottom. The Activated 
Carbon filter should be installed in the inner most filter slot.  

Note:  the  Prefilter  should  be  installed  before  the  Activated 
Carbon filter.  

5.1.3.3

 

Maintenance interval 

The replacement frequency for the Activated Carbon filter can 
be found in Table 1. 

5.1.3.4

 

Replacement Instructions  

To  replace  the  Activated  Carbon  filters  when  they  are  dirty, 
they  need  to  be  removed.  To  remove  the  Carbon  filters, 

remove the lower. Pull the Carbon filter toward you and out of 
the unit.   

The installation of new filters is the same as listed above. 

 

Содержание Air Medic TIF-06

Страница 1: ...IOM 2020 0 2360 Millrace Court Tel 905 670 3595 Mississauga Ontario Fax 905 670 3592 Canada L5N 1W2 Web Temspec com Air Medic Air Filtration Sterilization Unit TIF 06 Installation Operation Maintenance Manual ...

Страница 2: ...nel 3 4 4 Motor Access Panel 4 5 Air Medic Component Installation Maintenance and Replacement Instructions 4 5 1 Filters 4 5 1 1 Prefilter 4 5 1 2 HEPA Filter 5 5 1 3 Activated Carbon Filter Optional Accessory 5 5 2 Casters Optional Accessory 6 5 3 UVC Light Optional Accessory 6 5 4 Ionizer Optional Accessory 7 5 5 Fan Assembly 8 5 6 Differential Pressure Switch Optional Accessory 8 6 Warranty Inf...

Страница 3: ...e hot and could cause serious burns if not handled properly Please wait until the lamp has cooled to room temperature to remove from the unit WARNING UV Light Hazard Harmful to skin and eyes Can cause temporary or permanent loss of vision Never look at the lamp while illuminated To prevent exposure to ultraviolet light be sure the power is disconnected before servicing CAUTION Ionizer generates hi...

Страница 4: ... switched This pressure switch measures the air pressure around the HEPA filter and when it is too high the indicator light will come on to warn the user it should be changed This is only a warning indicator The unit will continue to work with it on but the airflow will be decreased due to the dirty filter 3 3 Fan Control The fan control is available in either variable speed or constant speed conf...

Страница 5: ...el is equipped with two cam latches that turn to lock in place to secure the panel and the optional lock mentioned above To open the lower access panel unlock the cam lock turn the cam latches to the unlocked position and then pull up and out When replacing the panel align the panel with the width of the unit and slide it down onto the hook at the bottom Make sure not to damage the power cable 4 3...

Страница 6: ...initial dust and dirt particles before they can get inside of the Air Medic similar to your HVAC equipment 5 1 1 2 Replacement Frequency The replacement frequency for the Prefilters can be found in Table 1 5 1 1 3 Replacement Instructions To replace the Prefilters when they are dirty they need to be removed To remove the Prefilters the lower access panel must be removed Lift out of the lower guide...

Страница 7: ...d then tighten the wingnuts 5 1 3 Activated Carbon Filter Optional Accessory 5 1 3 1 Description An optional accessory for the Air Medic is two 1 inch Activated Carbon Filters These filters make use of Activated Carbon to adsorb airborne particles to help with allergens and odours 5 1 3 2 Installation Instructions The Activated Carbon filters are installed on the intakes behind the Prefilters To i...

Страница 8: ...on the fan assemblies 5 3 1 2 Installation Instructions First remove the lower and upper access panel and then remove the HEPA filter Take the light ballast and align the tabs with the holes on the right fan bulkhead closest to the wall of the unit and screw them down with the provided Philips head screws Take supplied UVC mount assembly and align it with the four holes two on each bulkhead and wi...

Страница 9: ...ews Attach the ionizer quick connect to the 2 pin quick connect to the unit quick connect below the electrical cover Ensure they have a firm and properly sealed connection and that the locks have fully engaged 5 4 1 3 Maintenance Interval It is recommended that every six 6 months the ionizer being inspected for dust or dirt buildup on the brushes or indicator light If there is dust or dirt it shou...

Страница 10: ...l pressure switch This switch measures the pressure around the HEPA filter with the use of 1 4 tubing to let the operator know when the HEPA needs to be replaced When the pressure differential around the HEPA rises above the recommended level the switch will close and it will light the amber indicator light on the unit face 5 6 1 2 Maintenance Interval The switch is fully sealed and does not need ...

Страница 11: ...vated Carbon Filter 16 x 18 x 1 MERV 7 FPC 16118 1 year UVC Light 16 Watt EUL 01516 1 year Table 2 Unit Specifications TIF 06 TIF 08 Performance 450 CFM 600 CFM Purge Mode 450 CFM 800 CFM Purge Mode Electrical 120 VAC 2 6 FLA 6ft 2m Power cord 120 VAC 3 4 FLA 6ft 2m Power cord Fan Motor 250mm Backward Inclined Impeller fan with ECM 20 100 fan modulation 600C FM 1 2 TSP 285 s 300pa 280mm Backward I...

Страница 12: ...EMC 09004 CONSTANT SPEED FAN CONTROLLER EMC 09003 DOOR INTERLOCK SWITCH ESX 08000 DIFFERENTIAL PRESSURE SWITCH ESD 22000 CASTER WHEELS LOCKING WCX 51608 CASTER WHEELS WCX 51608 LIFT HANDLE VHX 01004 KEYED CAM LOCK NXL 01008 FAN BFE 025017 3 4IN FAN SCREW NSL 00306 IONIZER NUT NNX 21024A UVC THUMB NUT NNT 00202 CAM LOCK NXL 01010 3 4IN SCREW NSH 10206 1 2IN SCREW NSL 00304 CASTER LOCK WASHER NWX 21...

Отзывы: