Connect ITAG
®
4 data logger to PC via USB connection, open online configuration portal
www.itag4.com
in internet browser, then select the device you are going to configure, input
configuration parameters, download configuration profile (.cfg format) and save it to the
logger created Removable Storage Device "
TempSenDisk
".
Press and hold
button for
3
seconds to
activate logger, "Hold" on LCD reminds user to keep current
button action, until
appears on downleft corner of LCD.
will be replaced by customized Profile ID
number on LCD after configuration, if Profile ID was
applied. Please note Profile ID is only applicable to
ITAG4 firmware version 3.0 or higher.
Connect logger to PC, open TempCentre software, move to Configure Device mode, enter
configuration parameters, and click Apply to complete configuration.
Please
DO NOT
remove logger from PC before you are prompted of configuration success.
ITAG
®
4
/ BIO / TH
Data Recording Indicator
(Version X3.0 or above)
1> Configuration
2> Start Recording
3> Stop Recording
©2019. All rights reserved, TempSen Electronics Company
T: +86 21 6768 5210 F: +86 21 6768 5232 E: [email protected] W: www.tempsen.com
4> Download Data
Connect ITAG
®
4 data logger to USB port, and copy PDF report from removable storage
device
TempSenDisk
to your computer.
Configuration file contains local Time Zone information, so please download new
configuration file to avoid time error when you are in different time zone.
Please
DO NOT
connect ITAG4 to PC if you do not plan to use it immediately, first time
connecting ITAG4 to PC would turn on power and start consuming battery energy.
Please note two years shelf life is only applicable to ITAG4 loggers not connected to PC
yet. Factory programmed ITAG4 logger comes with one (1) year shelf life.
After first time connected to PC (no programming performed yet),
appears on the bottom of LCD, reminding you power already turned on.
5> Battery & Shelf Life
À
Ready for Manual Start
À
Waiting for Auto Start
À
Configuration success
»
Waiting for recording begins. Start Delay (30 minutes)
À
Expiry date: Sept 30, 2016
PDF report created by ITAG
®
4 contains raw data, so please
DO NOT
perform
any "Save" or "Save as" actions within any PDF reader software. Otherwise,
content of PDF file will be reorganized by PDF reader software and cause raw
data embedded in PDF file unrecognizable.
Please
DO NOT
remove ITAG4 from USB port while it's still in process of
generating PDF report.
Quick Guide EN
QG_190101_EN REV. A
You could also download device using TempCentre software.
Please pay attention to EXP date (Expiry Date) printed
on the back side of each logger every time starting
recording, please
DO NOT
use the logger if the EXP
date is earlier than current date.
Qualified for Air Transportation Monitoring
ITAG
®
4 data logger contains one(1) CR2032 lithium metal battery, which passed the Part
III, SubSection 38.3 of the UN Manual of Tests and Criteria.
ITAG
®
4 data logger is also certified to meet requirements in General Requirements and
Section II of Packaging Instruction 970, IATA Dangerous Goods Regulations
.
ITAG
®
4 complies with DO160G Environmental Conditions and Test Procedures for Airborne
Equipment.
Please download TempCentre software from https://www.tempsen.com/software/tempcentre
Press and hold
button until
(Stop) appears on home screen to stop
recording. You could disable STOP button during configuration.
ITAG4 logger could not be stopped by pressing button within 30 minutes after recording
begins.
If logger was set with start delay, recording begins after start delay
time elapsed.
▼ LCD (After Configuration)
▼ Option 1: Online Configuration
▼ Option 2: Configuration via TempCentre Software
À
Profile ID