Connect ITAG

®

4 data logger to PC via USB connection, open online configuration portal 

www.itag4.com

 in internet browser, then select the device you are going to configure, input 

configuration parameters, download configuration profile (.cfg format) and save it to the 
logger created Removable Storage Device "

TempSenDisk

".

Press and hold 

 button for 

3

 seconds to 

activate logger, "Hold" on LCD reminds user  to keep current 
button action, until 

 

 appears on downleft corner of LCD. 

 will be replaced by customized Profile ID 

number on LCD after configuration, if Profile ID was 
applied. Please note Profile ID is only applicable to 
ITAG4 firmware version 3.0 or higher.

Connect logger to PC, open TempCentre software, move to Configure Device mode, enter 
configuration parameters, and click Apply to complete configuration.

Please 

DO NOT

 remove logger from PC before you are prompted of configuration success.

ITAG

®

4

 

 / BIO / TH

 

 

Data Recording Indicator

(Version X3.0 or above)

1> Configuration

2> Start Recording

3> Stop Recording

©2019. All rights reserved, TempSen Electronics Company

T: +86 21 6768 5210  F: +86 21 6768 5232  E: [email protected]  W: www.tempsen.com

4> Download Data

Connect ITAG

®

4 data logger to USB port, and copy PDF report from removable storage 

device 

TempSenDisk 

to your computer. 

Configuration file contains local Time Zone information, so please download new 
configuration file to avoid time error when you are in different time zone.

Please 

DO NOT

 connect ITAG4 to PC if you do not plan to use it immediately, first time 

connecting ITAG4 to PC would turn on power and start consuming battery energy.

Please note two years shelf life is only applicable to ITAG4 loggers not connected to PC 
yet. Factory programmed ITAG4 logger comes with one (1) year shelf life.

After first time connected to PC (no programming performed yet),   
appears on the bottom of LCD, reminding you power already turned on.

5> Battery & Shelf Life

 

À

Ready for Manual Start

 

À

Waiting for Auto Start

 

À

Configuration success

 

»

Waiting for recording begins. Start Delay (30 minutes)

 

À

Expiry date: Sept 30, 2016

PDF report created by ITAG

®

4 contains raw data, so please 

DO NOT 

perform 

any "Save" or "Save as" actions within any PDF reader software. Otherwise, 
content of PDF file will be reorganized by PDF reader software and cause raw 
data embedded in PDF file unrecognizable.
Please 

DO NOT

 remove ITAG4 from USB port while it's still in process of 

generating PDF report.

Quick Guide EN

QG_190101_EN REV. A

You could also download device using TempCentre software.

Please pay attention to EXP date (Expiry Date) printed 
on the back side of each logger every time starting 
recording, please 

DO NOT

 use the logger if the EXP 

date is earlier than current date.

Qualified  for Air Transportation Monitoring

    

ITAG

®

4 data logger contains one(1) CR2032 lithium metal battery, which passed the Part 

III, SubSection 38.3 of the UN Manual of Tests and Criteria. 

ITAG

®

4 data logger is also certified to meet requirements in General Requirements and 

Section II of Packaging Instruction 970, IATA Dangerous Goods Regulations

.

ITAG

®

4 complies with DO160G Environmental Conditions and Test Procedures for Airborne 

Equipment.

Please download TempCentre software from https://www.tempsen.com/software/tempcentre

Press and hold 

 button until  

  (Stop) appears on home screen to stop 

recording. You could disable STOP button during configuration. 

ITAG4 logger could not be stopped by pressing button within 30 minutes after recording 
begins.

If logger was set with start delay, recording begins after start delay 
time elapsed.

 

▼ LCD (After Configuration)

 

▼ Option 1: Online Configuration

 

▼ Option 2: Configuration via TempCentre Software

 

À

Profile ID

Отзывы: