915CL Cleaver

Tempo Communications Inc.

1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155

3

Converting the 915CL to a Fixed Clamp Cleaver

Place the fixed clamp mechanism into the 915CL and secure with the screw 

provided.
The 900 micron and 250 micron fiber adapters can be permanently left in the 

915FS fusion splicer. If desired one M2X5 screws can be used to secure each 

fiber adapter into the fiber adapter mounting area of the 915FS.

Fixed Clamp

Mounting Screw

Blade Replacement

1.  Lay down cleaver with scale numbers on blade facing up.
2.  Remove blade locking screw.

3.  Remove old blade and replace with new blade. Do not touch blade (very 

sharp!) Use tweezers to remove old blade and insert new blade.

Normal Use and Maintenance

 (con’t)

4.  Lock new blade at position “1”.

5.  Lay 0.7 mm pencil lead across both pressure pads.

  0.7 mm 

 Pencil 

 Lead

6.  Move slider back and forth and check if blade touches pencil lead.

• If blade touches pencil lead:  Perform a cleave and check cleave angle 

on 915FS.
If cleave is not good:  Check if blade is damaged or dirty at 

position “1”.

• If blade drags pencil lead all the way across:  Go to “Blade Height 

Adjustment”. 

• If blade does not touch pencil lead:  Rotate pencil lead ≈180°  

and try again.
If not sure:  Perform a cleave and check cleave angle on 915FS.
If cleave is not good:  Go to “Blade Height Adjustment”.