OTHER ZERO ADJUSTMENTS 

 

 

 

Make a Zero Adjustment when a 

series of lab results

 shows a consistent error in the 

same direction with system in automatic.  

 

 

    Consistency error 

greater

 than 0.5% adjust Thermacoil. 

 

    Consistency error 

less

 than 0.5% shift LRV/URV with the Hart. 

 

MAGIC

 is reading high, ADD the error to the LRV & URV. 

 

 

MAGIC

 is reading low, SUBTRACT the error from the LRV & URV. 

 

 

 

 

 
 

 

After the First Zero Adjustment, output adjustments should be one half the error 
shown between the labs and the meter. 

 

 

TROUBLE SHOOTING TIPS

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Constant error between 

 

MAGIC 

and labs 

Check coil voltage setting 

*

 

– if okay refer to 

Ouput Adjustments to change zero 

Varying error with 
changes in consistency 

Refer to Output Adjustments to change 
sensitivity with the Curve Number 

MAGIC

  output does 

not change 
 
 
 

*

 Voltage can vary 

    +/- .05 VDC.

 

  Verify that breather plug on damper has 

been removed 

  Verify that shipping brace has been 

removed 

  Examine area around torque arm and 

motor for objects obstructing free 
movement  

  Disconnect loop power for a few seconds to 

reset micro.  Check power supply for a min 
of 16VDC @ 20mA output 

  Verify that the Box and Coil are working by 

pushing the torque arm to simulate the full 
range of output.   

   

       

 

LRV and URV Change EXAMPLE 

MAGIC

 = 3.5 % Con.; Labs Av. = 3.7 % Con.; Error = -0.2%  

MAGIC

 is reading low, so subtract error from LRV and URV 

LRV

 = 2 % - Adjust to 1.8%      

URV

 = 5 % - Adjust to 4.8% 

Output at the Control Room will go up .2% to a 3.7% Con. output. 

 
 

Содержание MAGIC II

Страница 1: ...leg 5 Connect digital voltmeter to front panel test points With motor off adjust Coil to DC voltage marked on S N tag Magic box 6 Use HART Communicator to set the Upper Range Value URV and Lower Range Value LRV to match control rooms settings 7 TEMCO has preset the curve for your consistency application 8 Electronic dampening is factory preset Do Not Change 9 Switch HART to read Process Variable 1...

Страница 2: ...mmunicator and then go to the LRV and URV screen 2 Set the Lower Range Value LRV to match control room setting 3 Set the Upper Range Value URV to match control room setting FIRST TIME ZERO ADJUSTMENT 1 Put Consistency Loop in Automatic Control 2 Obtain a Lab Sample Record the MAGIC s signal at the time the sample is taken 3 Adjust Coil so the MAGIC s signal agrees with Lab Sample SN Tag w Coil Vol...

Страница 3: ...MAGIC II Training Guide 11 14 Page 3 MAGIC CURVE SELECTION Furnish MAGIC II Curves Unbleached Stock 10 OCC and Bleached 7 TMP and Newsprint 6 The curve is factory preset for your application ...

Страница 4: ... refer to Ouput Adjustments to change zero Varying error with changes in consistency Refer to Output Adjustments to change sensitivity with the Curve Number MAGIC output does not change Voltage can vary 05 VDC Verify that breather plug on damper has been removed Verify that shipping brace has been removed Examine area around torque arm and motor for objects obstructing free movement Disconnect loo...

Страница 5: ...late 20mA output ____VDC 20mA 5 Check Coil Voltage with MOTOR OFF ____VDC 6 Does Coil Voltage go UP when torque arm is pushed to the right and DOWN when pushed to the left ____ Yes ____ No 7 Does Consistency Output follow the coil voltage changes ____ Yes ____ No 8 Disconnect ThermaCoil from electronics enclosure Does Coil Voltage equal approximately 10vdc ____ Yes ____ No TEMCO Engineering Inc 16...

Отзывы: