mx7b_im rev E Page 

7

 of 

23

 

4.

 

Set Up 

 

4.1

 

Installation 

4.1.1

 

Each MX7B is designed to be used in conjunction with a 7B sensor. The 7B sensor can be 

mounted in any position however it is recommended to mount it away from any contaminants 

that might be in the system as over time this can affect the measurement accuracy. Mounting 

the 7B sensor close to the area where vacuum measurement is desired is preferred to increase 

the accuracy. Each 7B sensor has a key that only allows for the MX7B to mate with the sensor in 

the proper orientation. Rotate the MX7B until the correct alignment is obtained and the MX7B is 

able to slide onto the 7B sensor. Connect the power connector to the top of the unit and be sure 

to tighten the screws so that a firm connection is maintained. 

 

4.1.2

 

Try to avoid connecting the MX7B to the vacuum chamber by long or narrow piping as this may 

affect the accuracy and response time of the unit. Avoid mounting the unit near a heater within 

the chamber as this may affect the measurement accuracy. Excessive vibration of the unit may 

affect accuracy and decrease the life of the unit. Exposure to oils and other contaminants will 

decrease the accuracy and decrease the life of the sensor. Do not expose the unit to corrosive 

gases. 

 

4.2

 

Electrical Information 

4.2.1

 

Power connectors and adapters for use with the MX7B are available for purchase from Televac®. 

Operators wishing to create their own adapters should refer to the pin out of the MX7B below. 

Each connection labeled “Floating” has no connection and should remain so. 

 

Pin  Description 

 

Analog Output (Ground) 

RS-485 A(-) 

SP2 NC 

SP2 NO 

SP1 Open Drain 

Starter Output (5V) 

Starter Output (Ground) 

Supply Voltage 

Analog Output (0-10 V) 

10 

RS-485 B(+) 

11 

SP2 Com 

12 

SP1 Source 

13 

Sensor Voltage Control 

14 

Sensor Voltage Control 

15 

Supply Voltage (Ground) 

 

4.3

 

Turning the High Voltage On and Off 

4.3.1

 

The MX7B has an operating range that does not include measurement above 1*10

3

 Torr. If the 

unit is operated above 1*10

3

 Torr, the sensor will be damaged and electrical arcing inside the 

sensor may occur. To prevent this, the unit should be paired with a low vacuum gauge, such as 

the MX4A or MX2A, to ensure safe operation. 

 

4.3.2

 

The MX7B provides two control methods for toggling the high voltage; Menu Control and Pin 

Control. To select which method is used, navigate to the Vsense Control menu screen and select 

either Menu or Pins. Pin Control is provided through pins 13 and 14 of the MX7B connector. 

 

 

Содержание MX7B

Страница 1: ...47 2500 mx7b_im rev E Page 1 of 23 MX7B Cold Cathode Ionization Gauge Instruction Manual Rev E 2022 Televac The Fredericks Company 2400 Philmont Avenue Huntingdon Valley PA 19006 web www frederickscompany com email sales frederickscompany com tel 1 215 947 2500 fax 1 215 947 7464 ...

Страница 2: ...ng the High Voltage On and Off 7 5 Menu Navigation 8 5 1 Description 8 5 2 Menu Structure 9 6 Explanation of Menu Items and Navigation 10 6 1 Measurement 10 6 2 High Voltage Toggle 10 6 3 System Settings 10 6 4 Calibration 12 6 5 Set Points 13 6 6 Output 13 7 Analog Output 14 7 1 Description 14 7 2 Linear 14 7 3 Linear by Decade 14 7 4 Logarithmic 15 8 Cleaning the MX7B Sensor 16 8 1 Cleaning Inst...

Страница 3: ...be changed through RS 485 communications with a PC through USB communications with a PC or through the unit s front panel using four capacitive touch buttons located on the top of the unit The MX7B has 2 set points one relay and one open collector and a selectable analog output option The analog output is a configurable 0 10 V output that includes options for linear linear by decade and logarithmi...

Страница 4: ... 2022 Televac The Fredericks Company www televac com sales televac com 1 215 947 2500 mx7b_im rev E Page 4 of 23 1 4 Dimensions ...

Страница 5: ...use of the instrument Televac disclaims all liability for the customer s failure to comply with these requirements 2 1 3 1 READ Instructions Read all safety and operating instructions before operating the product 2 1 3 2 RETAIN instructions Retain the safety and operating instructions for future reference 2 1 3 3 HEED warnings Adhere to all warnings on the product and in the operating instructions...

Страница 6: ...ety Instructions 3 2 1 Review all safety instructions outlined in Section 2 3 3 Connect Power 3 3 1 Connect a 24V power supply to the unit See the pin out in Section 4 2 for more information 3 4 Check the Measurement Screen 3 4 1 The unit should power up and display OFF on the measurement screen Do not attempt to turn on the high voltage 3 5 Power down the MX7B 3 5 1 Power down the MX7B and instal...

Страница 7: ...fe of the unit Exposure to oils and other contaminants will decrease the accuracy and decrease the life of the sensor Do not expose the unit to corrosive gases 4 2 Electrical Information 4 2 1 Power connectors and adapters for use with the MX7B are available for purchase from Televac Operators wishing to create their own adapters should refer to the pin out of the MX7B below Each connection labele...

Страница 8: ...Televac MX2A or MX4A to automatically control the unit via an open collector or relay For more information see Application Note AN 3023 4 3 5 To protect the sensor from damage if the sensor high voltage is toggled on when in either Menu Control or Pin Control mode the pressure is above 1 10 3 Torr overcurrent protection will turn the sensor off If the sensor is in Menu Control mode then a command ...

Страница 9: ...w Resolution 5 2 3 3 2 High Resolution 5 2 3 4 Vsense Control 5 2 3 4 1 Menu Control 5 2 3 4 2 Pin Control 5 2 3 5 Diagnostics 5 2 3 5 1 FW Date 5 2 3 5 2 FW Version 5 2 4 Calibration 5 2 4 1 1 10 6 Torr 5 2 4 2 1 10 5 Torr 5 2 4 3 1 10 4 Torr 5 2 4 4 1 10 3 Torr 5 2 4 5 Analog Output Adjustment 5 2 5 Set Points 5 2 5 1 SP1 Off 5 2 5 2 SP1 On 5 2 5 3 SP2 Off 5 2 5 4 SP2 On 5 2 6 Output 5 2 6 1 Lin...

Страница 10: ... Control Mode then the user must turn on the high voltage by sending a command over communications or by navigating to the High Voltage Toggle menu screen Once at the High Voltage Toggle screen press ENT to unlock the screen press UP or DOWN to change the setting and then press ENT to save the change The high voltage will then turn on and a pressure indication will be displayed on the Measurement ...

Страница 11: ...ure reading Gas Type Conversion Factor Nitrogen 1 00 Hydrogen 3 00 Helium 8 90 Neon 4 30 Argon 0 84 Krypton 0 59 Carbon Dioxide 0 69 6 3 4 Resolution 6 3 4 1 The MX7B displays two significant digits by default low resolution The user may choose to operate the gauge in high resolution mode in which the unit will display three significant digits 6 3 5 High Voltage Control 6 3 5 1 The MX7B has an upp...

Страница 12: ...hanging calibration settings voids NIST Traceable calibrations 6 4 4 10 6 Torr 6 4 4 1 The 1 10 6 calibration point allows the gauge to adjust the pressure reading at 1 10 6 Torr This calibration point should not be used to adjust the gauge reading below a reference pressure of 5 10 7 Torr or above 5 10 6 Torr 6 4 5 10 5 Torr 6 4 5 1 The 1 10 5 calibration point allows the gauge to adjust the pres...

Страница 13: ...ired reading is reached press ENT to save the change or SEL to navigate away without saving the change 6 6 Output 6 6 1 The MX7B provides a user selectable analog output as well as RS 485 and USB digital outputs 6 6 2 To modify the output settings 6 6 2 1 Navigate to the measurement screen 6 6 2 2 Press SEL five times 6 6 2 3 Press UP and DOWN to navigate between the analog output and RS 485 scree...

Страница 14: ...wable but not settable Use the high value 1 0 10H exponent H in the following equation to convert the voltage output to pressure 𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉 𝑂𝑂𝑂𝑂𝑂𝑂𝑂𝑂𝑂𝑂𝑂𝑂 10𝐻𝐻 1 𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃 7 2 2 To select the Linear output navigate to the Analog Output screen press ENT and then use the UP or DOWN arrow to select Linear Press ENT to save the change Then press DOWN once to navigate to the Output High screen...

Страница 15: ...𝑂𝑂𝑂𝑂𝑂𝑂𝑂 0 𝐿𝐿 𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃 𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇 10 𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉 𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 𝑜𝑜𝑜𝑜𝑜𝑜𝑜𝑜𝑜𝑜𝑜𝑜 7 4 2 For example with the logarithmic range set to a 0 V value of 1 10 7 and a 10 V value of 1 10 3 and a voltage reading of 3 075 Volts the pressure corresponds to 𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 10 3 7 2 5 𝑂𝑂𝑂𝑂𝑂𝑂𝑂𝑂𝑂𝑂𝑂𝑂 0 7 7 𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃 𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇 10 3 075 2 5 7 1 7 10 6 𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇 7 4 3 To select the Logarithmic output navigate to the Analog O...

Страница 16: ...t and the anode assembly should be replaced See below for relevant part numbers 8 1 7 Blast the vacuum wetted portions of the body with glass beads labeled with blue in the diagram concentrating on the pole piece and O ring areas but making sure to clean all surfaces wetted to vacuum If the threaded end of the tube needs to be cleaned it can be bead blasted as well 8 1 8 After cleaning use compres...

Страница 17: ...Communications Specifications heading for information on what settings are necessary to properly communicate with the MX7B 9 2 Changing Communications Settings 9 2 1 The communication parameters baud rate address etc are changed through the local menu Please see Section 6 in this manual entitled Explanation of Menu Items and Navigation for more information on navigating and editing within the menu...

Страница 18: ...d Response Description R1 xx Read units where xx is PA Pa TR Torr MB mbar R3 xx Read resolution where xx is LO low HI high R5 xyzzabb Read analog output range where x 1 linear 2 log 3 linear by decade y high value exponent sign zz high value exponent a low value exponent sign bb low value exponent R6 xx Read gas type where xx is N2 Nitrogen H2 Hydrogen NE Neon C2 Carbon Dioxide AR Argon HE Helium ...

Страница 19: ...thmic yzz high value abb low value y high value exponent sign zz high value exponent a low value exponent sign bb low value exponent W53 3 Set analog output format to linear by decade W6xx xx Set gas type where xx is N2 Nitrogen AR Argon H2 Hydrogen HE Helium NE Neon KR Krypton C2 Carbon Dioxide Response is new gas type W7YppsePPSE ppseePPSEE Set thresholds for set point Y where ppsee SPL PPSEE SP...

Страница 20: ...mic yzz high value abb low value y high value exponent sign zz high value exponent a low value exponent sign bb low value exponent W53 3 Set analog output format to linear by decade W6xx xx Set gas type where xx is N2 Nitrogen AR Argon H2 Hydrogen HE Helium NE Neon KR Krypton C2 Carbon Dioxide Response is new gas type W7YppsePPSE ppseePPSEE Set thresholds for set point Y where ppsee SPL PPSEE SPH ...

Страница 21: ...ion adjustment response is new pressure reading WC3xyy ppsee Set 1E 4 calibration adjustment response is new pressure reading WC4xyy ppsee Set 1E 3 calibration adjustment response is new pressure reading WC5xyy ppsee Write analog output adjustment Return value is pressure reading 9 4 6 USB Commands for Setting RS 485 Communications Settings Command Response Description C1 xxxx Return serial baud r...

Страница 22: ...EE 1E 4 Torr Adjustment is Baaa reading is PPSEE 0WC4Baaa CR PPSEE 1E 3 Torr Adjustment is Baaa reading is PPSEE 9 6 USB Sample Commands 9 6 1 Note that all USB commands are the same as the RS 485 commands with the star and address prefix omitted Command Output Description R1 CR 0001 Units are in Pa 0002 Units in Torr 0003 Units in mbar R2 CR ppseePPSEE Lower set point 1 given by ppsee Upper set p...

Страница 23: ...lf Command error an invalid character was sent for the command or an invalid number after the character 0N0002 cr lf Units error an invalid parameter was sent after a command specifier 9 8 Firmware Updates 9 8 1 The MX7B has the ability to update the firmware when in the field over the USB communications For more information on updating the firmware please contact sales frederickscompany com ...

Отзывы: