8
Maintenance
No maintenance is required beyond periodic replacement of a contaminated sensor. To replace the
sensor, remove the Philips head screw in the center of the instrument, then gently rock and pull off the
electronic assembly. Replace the old sensor with a new one, remount the electronics and replace the
Phillips head screw.
DO NOT OVER TIGHTEN THE SCREW!
CHAPTER 6
TELEVAC
CHAPTER 7
Calibration
All instrumentation is calibrated at the factory. No further calibration at the customer’s facility should be
required, even if the gauge tube is replaced.
For a reliable recalibration at a later date, it is necessary to have a reference standard vacuum system
whose pressure is known to be accurate. For such a calibration, the instrument, sensor tube and cable
should be returned to the factory for re-certification. However, if the vacuum calibration verification must
be done at the customer’s facility, a knowledgeable electronics technician trained in vacuum calibration
should operate equipment of certifiable accuracy.
CHAPTER 8
Troubleshooting
1. If display consistently displays “HI”, the pressure maybe above 20 20,000 microns. If a broken filament
is suspected, check for continuity between pins # 1 & # 8 and # 2 & # 7 on the sensor.
NOTE: If, after doing some or all of the above, the units are still inoperative, return to the factory for
repair.