CHAPTER 4
(cont.)
TELEVAC
To change the gas correction factor:
1) Assure the rear mounted operator
lock out switch
is in the
“OFF”
position
(the switch is in the down position)
2) Press the
SETUP
button once
3) The Nitrogen (N2) or Argon (Ar) display will flash
4) Press the
UP
or
DOWN
arrow buttons to select a gas correction, The flashing
display is the gas selected)
5) Press the
TEST OP
button to complete the selection
(
Note:
if the
TEST OP
switch is not pressed in 60 seconds the instrument will return to
normal operation and the change in gas correction is not accepted)
C
HANGE
M
EASUREMENT
U
NITS
:
The MC300 has the ability to display vacuum values in Pascals, MBAR or MICRONS
and TORR.
1) Assure the rear mounted operator
lock out switch
is in
“OFF”
position, (the
switch is in the down position)
2) Press the
SETUP
button once
3) Nitrogen (N2) and Argon (Ar) display will flash
4) Press the
SETUP
a second time
5) The Pascal, MBAR or TORR lamp will flash
6) Press the
UP
or
DOWN
arrow buttons to select the desired unit of measurement
7) Press the
TEST OP
button to complete the selection
(
Note:
if the
TEST OP
switch is not pressed in 60 seconds the instrument will return to
normal operation and the change in gas correction is not accepted)
A
SSIGN
S
ET
P
OINT
R
ELAYS AND
V
ALUES
:
The instrument has four relays that are assignable to any of the vacuum measurement
channels. Thermocouple (or Convection) Channel 1 is Channel 1 (CH1), thermocouple
(or Convection) Channel 2 is Channel 2 (CH2) and the Ion gauge is Channel 3 (CH3).
Setpoint SP 1 is relay #1, setpoint SP 2 is relay #2 etc.