background image

 

Page 

2

 of 

3

 

 

should be removed. Do not use the airway if the blockage or 
particle cannot be removed. 

• Deflate the cuff completely. Once deflated, check the cuff for 

spontaneous inflation. Do not use the airway if the cuff 
spontaneously inflates. 

 

9. DEFLATING THE DEVICE PRIOR TO INSERTION 

-  After firmly connecting a syringe of at least 50ml to the inflation port, 

hold the syringe and the LMA Supreme™ exactly as shown in figure 4. 
Move the connected syringe away from the device until the inflation 
line is slightly stretched as shown. Compress the distal end of the 
device in between the index finger and thumb while withdrawing air 
until a vacuum has been obtained. 

- While deflating, hold the device so that the distal end is curled slightly 

anteriorly as shown in Figure 4 

- Deflate the device until the tension in the syringe indicates a vacuum 

has been created in the mask. Keep the syringe under tension while 
rapidly disconnecting it from the inflation port. This will ensure the 
mask remains correctly deflated, as shown in Fig 5.         

                                                   

             

 

Figure 4:

 LMA Supreme™ deflation       

Figure 5:

 After achieving wedge 

shape cuff during deflation, 
disconnect the syringe from the 
inflation line  

 

10. INSERTION 

-  Lubricate the posterior surface of the mask and airway tube just prior 

to insertion. 

-  Stand behind or beside patient’s head. 
-  Place the head in the neutral or slight “sniffing” position (Sniffing = 

extension of head + flexion of neck). 

-  Hold the device exactly as shown in figure 6. 
-  Press the distal tip against the inner aspect of the upper teeth or 

gums.  

-  Slide inwards using a slightly diagonal approach (direct the tip away 

from the mid-line). 

-  Continue to slide inwards rotating the hand in a circular motion so 

that the device follows the curvature behind the tongue. 

-  Resistance should be felt when the distal end of the device meets the 

upper esophageal sphincter. The device is now fully inserted

 

 
 

 

Figure 6:

 Press the tip of the                     

Figure 7:

 Press the cuff further 

mask against the hard palate.                   into the mouth, maintaining 

 

  pressure against the palate

.

 

 

Figure 8:

 Swing the device inward             

Figure 9:

 Advance the device 

with a circular motion, pressing                 into the hypopharynx until 
against the contours of the hard                resistance is felt. 
and the soft palate.  

 

 

11. FIXATION 

Secure the LMA Supreme™ to patient’s face using adhesive tape as 
follows:  
-  Use a piece of adhesive tape 30-40cm long, holding it horizontally by 

both ends. 

-  Press the adhesive tape transversely across the fixation tab, 

continuing to press downwards so that the ends of the tape adhere 
to each of the patient’s cheeks and the device itself is gently pressed 
inwards by the tape.  

-  Do not rotate the tape around the proximal end of the device.  
-  Do not use a Guedel airway; the device has an integral bite block.  

                              

                                  

 

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           

                                 

     

Figure 10a  

Figure 10b

          

 

 

Figure 10:

 Fix the device in place using adhesive tape. Stretching the 

adhesive tape vertically downwards (See Figure 10a) ensures that the 
middle of the tape is pressed vertically downwards over the tab as 
shown in Figure 10b.

 

 

12. INFLATION  

Inflate the cuff with air until the appropriate intra-cuff pressure is 
reached. The recommended intra-cuff pressure should never exceed 
60cm H₂O. If no manometer is by hand, inflate with just enough air to 
achieve a seal sufficient to permit ventilation without leaks. 

 

Airway 
Size

 

Patient 
Weight

 

Max 
Size OG 
Tube

 

Recommended 
Maximum 
Inflation  
Volume 

 

Optimum 
Intra-Cuff  
Pressure

 

< 5kg

 

6Fr 

5 ml 

60cm H₂O 

1.5 

5-10kg 

6Fr 

8ml 

10-20kg

 

10Fr 

12 ml 

2.5 

20-30kg 

10Fr 

20ml 

30-50kg

 

14Fr 

30 ml 

50-70kg

 

14Fr 

45 ml 

70-100kg

 

14Fr 

45 ml 

Table 1:

 LMA Supreme™ selection guide  

 

13. CORRECT POSITION  

Correct placement should produce a leak-free seal against the glottis 
with the mask tip at the upper esophageal sphincter. The integral bite 
block should lie between the teeth. 

 

To facilitate diagnosis of correct mask placement, place a small bolus 
(1-2ml) of suitably viscous water soluble lubricant in the proximal end 
of the drain tube. In a properly placed mask, there should be a slight 
up-down meniscus movement of the lubricant following the 
application and release of gentle pressure on the suprasternal notch. 
This indicates that the distal end of the drain tube is correctly placed so 
that it seals around the upper esophageal sphincter (the ‘suprasternal 
notch test’). A similar movement may also be seen when gentle manual 
positive pressure is applied to the airway through the device. 

 

14. GASTRIC DRAINAGE  

The drain tube facilitates channeling of fluids and gases emerging from 
the stomach. To facilitate gastric drainage, a gastric tube may be 
passed through the drain tube into the stomach at any time during the 
anesthetic procedure. Refer to Table 1 for maximum gastric tube sizes. 
The gastric tube should be well lubricated and passed slowly and 
carefully. Suction should not be performed until the gastric tube has 
reached the stomach. Suction should not be applied directly to the end 
of the drain tube, as this may cause the drain tube to collapse and 
might theoretically cause injury to the upper esophageal sphincter.  

 

15. ANESTHESIA MAINTENANCE  

The LMA Supreme™ is well tolerated in spontaneously breathing 
patients when used with volatile agents or intravenous anesthesia, 
provided anesthesia is adequate to match the level of surgical stimulus 
and the cuff is not over-inflated. 

 

During PPV using the LMA Supreme™ tidal volumes should not exceed 
8ml/kg and peak inspiratory pressures should be kept below the 
maximum airway seal pressure. 

 

If leaks occur during PPV, this may be due to light anaesthesia causing a 
degree of glottic closure, severe reduction in lung compliance related 
to the procedure or patient factors or displacement or migration of the 
cuff by head turning or traction in an inadequately fixed mask. 

 

 

16. RECOVERY 

Removal should always be carried out by trained personnel. Although 
the device may not be removed in the operating theatre, its low 
invasivity makes it a good device to maintain the airway during 
recovery in the Post Anesthetic Care Unit (PACU) provided staff are 
appropriately trained and equipped. Because recovery involves 
increase in pharyngeal tone, it makes sense to reduce the volume of air 
in the cuff before sending the patient to the PACU; however, the cuff 
must never be fully deflated at this point. 

 

Fully deflate the cuff and simultaneously remove the device ONLY 
when the patient can open the mouth on command. If the cuff is FULLY 
deflated before the return of effective swallowing and cough reflexes, 
secretions in the upper pharynx may enter the larynx, provoking 
coughing or laryngeal spasm. 

 

Patient monitoring should continue throughout the recovery stage. 
Where appropriate, oxygen may be continuously administered through 
the anaesthetic circuit or via a T-piece attached to the proximal end of 
the airway device.

 

 

17. USE WITH MAGNETIC RESONANCE IMAGING (MRI)  

 

 

Testing has been performed to determine the compatibility of LMA 
Supreme™ with MRI. Prior to using LMA Supreme™ in this environment, 
the user should carefully compare the equipment and test conditions 
described with those planned for use in the actual clinical environment. 
Refer to the below for detailed results of device testing in the MRI 
environment. 

 

The 

LMA Supreme™ 

was determined to be MR-conditional. Non-

clinical testing demonstrated that the LMA Supreme™ is MR 
Conditional. A patient with this device can be scanned safely 
immediately after placement under the following conditions: 

 

Static Magnetic Field 

-Static magnetic field of 3-Tesla or less 
-Maximum spatial gradient magnetic field of 720-Gauss/cm or less 

 

MRI-Related Heating 

In non-clinical testing, the device produced the following temperature 
rise during MRI performed for 15-min of scanning (i.e., per pulse 
sequence) in the 3-Tesla (3-Tesla/128-MHz, Excite, HDx, Software 
14X.M5, General Electric Healthcare, Milwaukee, WI) MR system: 

 

Highest temperature 1.6°C 

 

Therefore, the MRI-related heating experiments for the device at 3-
Tesla using a transmit/receive RF body coil at an MR system reported 
whole body averaged SAR of 2.9 -W/kg (i.e., associated with a 
calorimetry measured whole body averaged value of 2.7-W/kg) 
indicated that the greatest amount of heating that occurred in 
association with these specific conditions was equal to or less than 
+1.6°C.

 

 
 

Artifact Information 

MR image quality may be compromised if the area of interest is in the 
exact same area or relatively close to the position of the device. 
Therefore, optimization of MR imaging parameters to compensate for 
the presence of this device may be necessary.  

 

Pulse 
Sequence 

T1-SE 

T1-SE 

GRE 

GRE 

Signal Void 
Size 

5,481-
mm

3,400-mm

12,343-
mm

7,394-mm

Plane 
Orientation 

Parallel 

Perpendicular 

Parallel 

Perpendicular 

 
18. SYMBOLS DEFINITION  

 

Manufacturer 

 

Authorized representative in the 
European Community 

 

Consult IFU on this website: 
www.LMACO.com 

 

Air inflation volume 

 

Patient weight 

 

 

 

Read Instruction before use 

 

 

 

Not made with natural rubber latex 

 

 

 

Fragile, handle with care 

 

 

 

Keep away from sunlight 

 

 

 

Keep dry 

 

 

 

This way up 

 

Product Code 

 

Lot Number 

 

 

 

CE Mark 

 

Do not Re-use 

 

 

 

Contains or Presence of Phthalates: 
Bis(2-ethylhexyl) phthalate (DEHP) 

 

 

 

Sterilized by Ethylene Oxide 

 

 

 

Use By 

 

Do not use if package is damaged 

 

 

 

Отзывы: