7
SENSITIVITY
0.002, 0.005, 0.01,0.02,
0.05, 0.1, 0.2, 0.5, 1.0,
2.0, set lamp and optics
Rotary selector switch
Used to establish the full-scale absorbance range.
Set lamp and optics: This position overrides AUTO
BASELINE and provides equal voltage to the opti-
cal unit photocells. At this time, the optical unit
should be balanced. It also initializes the lamp cur-
rent adjust circuit. (Type 11 only).
8
RECORDER OFFSET
Rotary control
Used for fine positioning of the absorbance
baseline (zero).
9
NOISE FILTER
0.5, 1.5, 5.0
3 position selector switch Reduces noise by employing a Bessel-type active
filter. The 0.5, 1.5 and 5.0 positions indicate the
approximate theoretical response time in seconds
for 10 to 90 percent full-scale output deflection.
10
Standby-Operate LED
Indicator lamp
Lights to indicate that secondary power is applied
to circuitry and the unit is in operational status.
RED: Type 11 lamp is warming up. Red after warm
up indicates lamp current needs to be reset!
GREEN: Type 11 lamp current is proper, ready for
operation.
Table 1-2 Front Panel Controls and Indicators (Continued)
Item in
Fig. 1-2
Control or Indicator
Description
Function
11
10
Содержание UA-6
Страница 2: ......
Страница 4: ......
Страница 12: ...x ...
Страница 16: ...Isco UA 6 Table of Contents xiv ...
Страница 32: ...UA 6 Detector Section 2 Preparation for Use 2 8 ...
Страница 41: ...UA 6 Detector Section 3 Operation 3 9 Figure 3 6 Ferrule Application Information ...
Страница 47: ...UA 6 Detector Section 3 Operation 3 15 Figure 3 9 Absorbance Curves of Four Commonly Used Solvents ...
Страница 52: ...UA 6 Detector Section 3 Operation 3 20 Figure 3 14 Problem Analysis Drift and Noise ...
Страница 54: ...UA 6 Detector Section 3 Operation 3 22 ...
Страница 65: ...UA 6 Detector Section 4 Theory of Operation 4 11 Figure 4 5 Type 11 Absorbance Detector Block Diagram ...
Страница 66: ...UA 6 Detector Section 4 Theory of Operation 4 12 ...
Страница 90: ...UA 6 Detector Appendix A Replacement Parts A 2 ...
Страница 91: ...UA 6 Detector Appendix A Replacement Parts A 3 ...
Страница 92: ...UA 6 Detector Appendix A Replacement Parts A 4 MAIN CIRCUIT BOARD ...
Страница 93: ...UA 6 Detector Appendix A Replacement Parts A 5 ...
Страница 94: ...UA 6 Detector Appendix A Replacement Parts A 6 ...
Страница 95: ...UA 6 Detector Appendix A Replacement Parts A 7 ...
Страница 96: ...UA 6 Detector Appendix A Replacement Parts A 8 ...
Страница 97: ...UA 6 Detector Appendix A Replacement Parts A 9 ...
Страница 98: ...UA 6 Detector Appendix A Replacement Parts A 10 ...
Страница 99: ...UA 6 Detector Appendix A Replacement Parts A 11 ...
Страница 100: ...UA 6 Detector Appendix A Replacement Parts A 12 ...
Страница 101: ...UA 6 Detector Appendix A Replacement Parts A 13 ...
Страница 102: ...UA 6 Detector Appendix A Replacement Parts A 14 ...
Страница 103: ...UA 6 Detector Appendix A Replacement Parts A 15 ...
Страница 104: ...UA 6 Detector Appendix A Replacement Parts A 16 ...
Страница 105: ...UA 6 Detector Appendix A Replacement Parts A 17 ...
Страница 106: ...UA 6 Detector Appendix A Replacement Parts A 18 This page intentionally left blank ...
Страница 107: ...UA 6 Detector Appendix A Replacement Parts A 19 ...
Страница 108: ...UA 6 Detector Appendix A Replacement Parts A 20 ...
Страница 109: ...UA 6 Detector Appendix A Replacement Parts A 21 Drawing modified 7 6 11 to remove obsolete flow cells ...
Страница 110: ...UA 6 Detector Appendix A Replacement Parts A 22 ...
Страница 111: ...UA 6 Detector Appendix A Replacement Parts A 23 ...
Страница 112: ...UA 6 Detector Appendix A Replacement Parts A 24 ...
Страница 114: ...UA 6 Detector Index Index 2 Lamp Replacement 5 11 Theory of Operation 4 8 Type 12 Fuses 2 4 ...
Страница 116: ......
Страница 118: ......
Страница 120: ...Warranty ...
Страница 122: ......