08354B DCN7877
Teledyne API T640 PM Mass Monitor
42
DOWNLOADING DAS (DATA ACQUISITION SYSTEM) DATA
DAS data (collected by the Datalogger, Section 3.2.7) can be downloaded from the
instrument to a flash drive through the Utilities>USB Utilities>DAS Download
menu, as follows:
1. Insert a flash drive into a front panel USB port and wait for the Status field to
indicate that the drive has been detected; available buttons will be enabled.
Figure 3-5:
Das download Page
2. To copy the data to the flash drive, press the Start button next to “Download
DAS Data from Instrument.” (The Cancel button will be enabled).
3. When complete, as indicated in the Status field, the Cancel button becomes
the Done button: press Done and then remove the flash drive.
UPDATING SOFTWARE/FIRMWARE
If the system is connected to a network that is connected to the Internet, it can check
for updates automatically or when the user prompts for a check. It is also possible
to check for updates or reload current software/firmware manually. This section
describes the various methods.
Automatic Update Notifications
When this feature is set to True (Setup>Vars>Periodically Check for Updates), the
system will check every week for an update and trigger an Alert when an update is
available, prompting the user to download the update