08354B DCN7877
Teledyne API T640 PM Mass Monitor
24
6. Once the ASC is leveled, set screws should be hand tightened until snug.
Do
not over-tighten; the screws should be relatively easy to loosen when
necessary
. Press slightly on instrument top cover to allow more clearance if
needed.
Do not use thread locker
.
7. Connect the ASC power and communications cable to the rear panel of the
T640 labeled ASC Connector (or IADS Connector in some cases).
8. Locate the 25’ temperature sensor cable (15’ in some cases). Plug the ambient
temperature probe connector into its respective rear panel electrical port.
Figure 2-12. Ambient Temperature Probe and ASC Connections (T640 with
640X Option shown)