Error Code
Description
-100 Command
Error
This is the generic syntax error for devices that
cannot detect more specific errors. This code
indicates only that a Command Error as defined in
IEEE 488.2,11.5.1.1.4 has occurred.
-102 Syntax error
An unrecognized command or data type was
encountered; for example, a string was received
when the device does not accept strings.
-103 Invalid
separator
The parser was expecting a separator and
encountered an illegal character; for example, the
semicolon was omitted after a program message
unit, MEAS:VOLT:DC?:MEASCURR:DC?
-104 Data type
error
The parser recognized a data element different
than one allowed; for example, numeric or string
data was expected but block data was
encountered.
-108 Parameter
not allowed
More parameters were received than expected for
the header; for example, the KLOCk command
only accepts one parameter, so receiving
SYSTem:KLOCk 1,0 is not allowed.
-109 Missing
parameter
Fewer parameters were recieved than required for
the header; for example, the KLOCk command
requires one parameter, so receiving KLOCk is not
allowed.
-111 Header
separator error
A character which is not a legal header separator
was encountered while parsing the header; for
example, no white space followed the header, thus
*SRE2 is an error.