Teledyne API – Model T300/T300M CO Analyzer
Communications Setup and Operation
151
6.5.2.
CONFIGURING ETHERNET COMMUNICATION USING DYNAMIC
HOST CONFIGURATION PROTOCOL (DHCP)
The default Ethernet setting is DHCP.
1. See your network administrator to affirm that your network server is running DHCP.
2. Access the Communications Menu (SETUP>MORE>COMM) and follow the setup
sequence as shown in Figure 6-5.
SETUP X.X COMMUNICATIONS MENU
ID
INET
COM1 COM2
EXIT
SETUP X.X ENTER PASSWORD:818
8
1
8
ENTR EXIT
SETUP X.X DHCP:ON
<SET SET>
EDIT
EXIT
SETUP X.X INST IP:0.0.0.0
<SET SET>
EXIT
SETUP X.X GATEWAY IP:0.0.0.0
<SET SET>
EXIT
SETUP X.X SUBNET MASK IP:0.0.0.0
<SET SET>
EXIT
SETUP X.X
TCP PORT:3000
<SET SET>
EDIT
EXIT
SETUP X.X ONLINE:ON
<SET SET>
EDIT
EXIT
SETUP X.X INITIALIZING INET 0%
d
EDIT button
is disabled
when DHCP
is ON.
DO NOT alter unless
instructed to by Teledyne
Instruments’ customer
Service personnel.
INITIALIZATION process proceeds
automatically
SETUP X.X COMMUNICATIONS MENU
ID ADDR
INET
EXIT
SETUP X.X INITIALIZATION SUCCEEDED
SETUP X.X INITIALIZATION FAILED
Contact your
IT Network
Administrator.
SETUP X.X
TCP PORT:3000
<SET SET>
EDIT
EXIT
SETUP X.X HOSTNAME: T300
<SET SET>
EDIT
EXIT
Both TCP ports
are inactive when
this is set for
OFF.
Figure 6-5 :
COMM – LAN / Internet Automatic Configuration (DHCP)
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...