Appendix C

 

Warranty/Repair Questionnaire 

T200U/UP and M200EU/EUP

 

(06863B, DCN6247) 

 

 

TELEDYNE API CUSTOMER SERVICE 

EMAIL: 

[email protected]

   

PHONE: (858) 657-9800 

TOLL FREE: (800) 324-5190 FAX: (858) 657-9816 

 

C-1 

CUSTOMER: ____________________________________ PHONE:  ___________________________________________ 

CONTACT NAME: _______________________________ FAX NO.  ___________________________________________ 

SITE ADDRESS: ______________________________________________________________________________________ 

MODEL SERIAL NO.: ____________________________ FIRMWARE REVISION: _______________________________ 

1.

 

ARE THERE ANY FAILURE MESSAGES?  ____________________________________________________________ 

_____________________________________________________________________________________________________ 

_____________________________________________________________________________________________________

 

 

PLEASE COMPLETE THE FOLLOWING TABLE: 

(NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST 

PARAMETERS SHOWN BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT

*IF OPTION IS INSTALLED 

PARAMETER RECORDED 

VALUE ACCEPTABLE 

VALUE 

RANGE 

PPB 

50 PPB TO 2,000 PPB 

NOx STAB 

PPB 

 0.1 PPB WITH ZERO AIR 

SAMPLE FLOW 

CM

1,000 ± 100 

OZONE FLOW 

CM

80 ± 15 

PMT SIGNAL WITH 

ZERO AIR 

MV -20 

TO 

150 

PMT SIGNAL AT SPAN 

GAS CONC 

MV 

PPB 

0-5,000MV 

0-2,000 PPB 

NORM PMT SIGNAL AT 

SPAN GAS CONC 

MV 

PPB 

0-5,000MV 

0-2,000PPB 

PREREACT 

MV -20 

TO 

150 

HVPS 

400 – 900 

RCELL TEMP 

ºC 

40 ± 1 

BOX TEMP 

ºC 

AMBIENT ± 5ºC  

PMT TEMP 

ºC 

5 ± 2ºC 

MF TEMP 

ºC 

40 ± 1ºC 

U/EU models only: 

MOLY TEMP 

ºC 

315 ± 5ºC 

UP/EUP models only: 

O3KL TEMP 

ºC 

200 ± 5ºC 

RCEL PRESS 

IN-HG-A <4” 

(CONSTANT) 

SAMP PRESS 

IN-HG-A 

~ 1-5” < AMBIENT (CONSTANT) 

NOx SLOPE 

 

1.0 ± 0.3 

NOx OFFSET 

 -50 

TO 

50 

NO SLOPE 

 

1.0 ± 0.3 

NO OFFSET 

 -50 

TO 

50 

ETEST 

PMT MV 

2000 ± 1000 

OTEST 

PMT MV 

2000 ± 1000 

 

Values are in the Signal 

I/O 

 

REF_4096_MV 

MV 

4096mv ±2mv and Must be Stable 

REF_GND 

MV 

0± 0.5 and Must be Stable 

 

06861C DCN6670

Содержание T200U

Страница 1: ...OLLUTION INSTRUMENTATION TAPI 9970 CARROLL CANYON ROAD SAN DIEGO CA 92131 1106 USA Toll free Phone 800 324 5190 Phone 1 858 657 9800 Fax 1 858 657 9816 Email api sales teledyne com Website http www teledyne api com Copyright 2010 2013 06861C DCN 6670 Teledyne Advanced Pollution Instrumentation 19 February 2013 ...

Страница 2: ......

Страница 3: ...addendum while the questionnaire includes Appendix C in its title there is no Appendix A or Appendix B with this addendum A Repair Questionnaire is inserted at the end of this addendum while the questionnaire includes Appendix C in its title there is no Appendix A or Appendix B with this addendum REVISION HISTORY Date Rev DCN Change Summary 2013 Feb 19 C 6670 Administrative and other minor updates...

Страница 4: ...ii This page intentionally left blank ...

Страница 5: ... Technician Symbol All operations marked with this symbol are to be performed by qualified maintenance personnel only Electrical Ground This symbol inside the instrument marks the central safety grounding point for the instrument CAUTION This instrument should only be used for the purpose and in the manner described in this manual If you use this instrument in a manner other than that for which it...

Страница 6: ...urface chaude Ne pas toucher Toucher à certaines parties de l instrument sans protection ou sans les outils appropriés pourrait entraîner des dommages aux pièces ou à l instrument Pictogramme technicien Toutes les opérations portant ce symbole doivent être effectuées uniquement par du personnel de maintenance qualifié Mise à la terre Ce symbole à l intérieur de l instrument détermine le point cent...

Страница 7: ...ations and Agency Approvals 9 2 1 EPA Equivalency Designation 10 2 2 Approvals and Certifications 11 2 2 1 SAFETY 11 2 2 2 EMC 11 2 2 3 OTHER TYPE CERTIFICATIONS 11 3 Instrument Layout 13 4 Pneumatics 15 5 Test functions 17 6 Spare Parts List 19 ...

Страница 8: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 vi This page intentionally left blank ...

Страница 9: ...is gold plated In addition there is a pre reactor which allows the measurements of any hydrocarbon interferents in the sample stream The sample flow is 1000 cc min The T200U is equipped with a high performance pump capable of producing a reaction cell pressure of less than 5 Inch Hg absolute Because the T200U is a close derivative of the T200 the Model T200 instruction manual continues to be valid...

Страница 10: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 8 This page intentionally left blank ...

Страница 11: ...er 24 hrs Lag Time1 20 seconds Rise Fall Time1 3 50 sec to 95 Linearity 1 of full scale or 0 1 ppb whichever is greater Precision 0 5 of reading above 5 ppb Sample Flow Rate 1100 cm3 min 10 Power Analyzer 100V 120V 60 Hz 62W 220V 240V 50 Hz 58W Power Ext Pump 110V 120V 60 Hz 240W 220V 240V 50 Hz 155W Analog Output Ranges All outputs 10V 5V 1V 0 1V selectable Three outputs convertible to 4 20 mA is...

Страница 12: ...ensions H x W x D 7 x 17 x 23 5 178mm x 432 mm x 597 mm Weight Analyzer 40 lbs 18 kg Weight Ext Pump Pack 21 lbs 9 5 kg Environmental4 Installation Category Over voltage Category II Pollution Degree 2 1 As defined by the USEPA 2 At constant temperature and voltage 3 With adaptive filter 20 ppb change 2 1 EPA Equivalency Designation Teledyne API s T200U nitrogen oxides analyzer is designated as a r...

Страница 13: ...ment for measurement control and laboratory use CE 2006 95 EC Low Voltage Directive North American cNEMKO Canada CAN CSA C22 2 No 61010 1 04 NEMKO CCL US UL No 61010 1 2nd Edition 2 2 2 EMC EN 61326 1 IEC 61326 1 Class A Emissions Industrial Immunity EN 55011 CISPR 11 Group 1 Class A Emissions FCC 47 CFR Part 15B Class A Emissions CE 2004 108 EC Electromagnetic Compatibility Directive 2 2 3 OTHER ...

Страница 14: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 12 This page intentionally left blank ...

Страница 15: ...13 3 INSTRUMENT LAYOUT Figure 3 1 shows the internal chassis layout of the T200U ...

Страница 16: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 14 Figure 3 1 T200U Internal Chassis Layout ...

Страница 17: ...ects of hydrocarbons in NOx measurements In addition the T200U is equipped with a high performance pump capable of producing a reaction cell pressure of less than 5 Inch Hg absolute SAMPLE PRESSURE SENSOR VACUUM PRESSURE SENSOR FLOW PRESSURE SENSOR PCA INSTRUMENT CHASSIS PERMAPURE DRYER PREREACTOR Orifice Dia 0 004 PMT GOLD REACTION CELL PUMP SAMPLE GAS INLET NO NOX VALVE NO2 Converter O3 Destruct...

Страница 18: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 16 This page intentionally left blank ...

Страница 19: ...17 5 TEST FUNCTIONS Some test functions differ between the T200U and the T200 Please refer to Appendix C this addendum for the test functions in the T200U ...

Страница 20: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 18 This page intentionally left blank ...

Страница 21: ...ares Stocking Levels for your analyzer Note Use of replacement parts other than those supplied by TAPI may result in non compliance with European standard EN 61010 1 Note Due to the dynamic nature of spare parts lists please refer to the Website or call Customer Service for most recent updates ...

Страница 22: ...LTER 47MM KB 016300900 ASSY SAMPLE FILTER 47MM ANG BKT 5UM 024600100 ASSY PREREACTOR NO NOX VALVE 024930000 CD PMT NOX 037860000 ORING TEFLON RETAINING RING 47MM KB 039700100 HEATER BAND TYPE K DUAL VOLTAGE KB 040010000 ASSY FAN REAR PANEL 040030800 PCA PRESS SENSORS 2X FLOW NOX 040400000 ASSY HEATERS THERMAL SWITCH RX CELL 040410500 ASSY VACUUM MANIFOLD 040420200 ASSY O3 GEN BRK HIGH O P 04151020...

Страница 23: ...EANSER ALUMINUM 059940000 OPTION SAMPLE GAS CONDITIONER NOX 062390000 ASSY MOLY GUTS w WOOL 062420200 PCA SER INTRFACE ICOP CPU E OPTION USE WITH ICOP CPU 062870000 064540000 ASSY PUMP NOX INTERNAL 115V 60HZ 064540100 ASSY PUMP NOX INTERNAL 230V 60HZ 064540200 ASSY PUMP NOX INTERNAL 230V 50HZ 066970000 PCA INTRF LCD TOUCH SCRN F P 067240000 CPU PC 104 VSX 6154E ICOP 067300000 PCA AUX I O BD ETHERN...

Страница 24: ...NG 2 006VT KB OR0000002 ORING 2 023V OR0000025 ORING 2 133V OR0000027 ORING 2 042V OR0000034 ORING 2 011V FT10 OR0000039 ORING 2 012V OR0000044 ORING 2 125V OR0000046 ORING 2 019V OR0000083 ORING 105M 1MM W X 5 MM ID VITON OR0000086 ORING 2 006 CV 75 COMPOUND KB OR0000094 ORING 2 228V 50 DURO VITON KB PU0000047 PUMP THOMAS 2688VE44 2 HEAD 115V 60HZ PU0000048 PUMP THOMAS 2688VGH144 230V 50 60HZ CE ...

Страница 25: ..._______________________________________________________ PLEASE COMPLETE THE FOLLOWING TABLE NOTE DEPENDING ON OPTIONS INSTALLED NOT ALL TEST PARAMETERS SHOWN BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT IF OPTION IS INSTALLED PARAMETER RECORDED VALUE ACCEPTABLE VALUE RANGE PPB 50 PPB TO 2 000 PPB NOx STAB PPB 0 1 PPB WITH ZERO AIR SAMPLE FLOW CM3 1 000 100 OZONE FLOW CM3 80 15 PMT SIGNAL WITH ZERO A...

Страница 26: ...T TEST S HAVE YOU DONE TRYING TO SOLVE THE PROBLEM ________________________________ ___________________________________________________________________________________________________ ___________________________________________________________________________________________________ ___________________________________________________________________________________________________ ________________...

Отзывы: