Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Operating Instructions
87
Table 4-5: Secondary Setup Mode Features and Functions
MODE OR FEATURE
CONTROL
BUTTON
DESCRIPTION
MANUAL
SECTION
External Communication
Channel Configuration
COMM
Used to set up and operate t
he analyzer’s various external I/O
channels including RS-232; RS 485, modem communication
and/or Ethernet access.
System Status Variables
VARS
Used to view various variables related to the instruments current
operational status
System Diagnostic Features
DIAG
Used to access a variety of functions that are used to configure,
tes
t or diagnose problems with a variety of the analyzer’s basic
systems
IMPORTANT
Any changes made to a variable during one of the following procedures are not
acknowledged by the instrument until the ENTR button is pressed. If the EXIT button is
pressed before the ENTR button, the analyzer will beep, alerting the user that the newly
entered value has not been accepted.
S
ETUP
M
ODE
P
ASSWORD
S
ECURITY
When the T100’s
SETUP
mode is activated, the instrument will prompt the user to enter
a security password. The default password is
818
. This allows access to all of the
instruments basic functions and operating modes as well as some of its more powerful
diagnostic tools and variables.
The analyzer will automatically insert
818
into the password prompt field. Simply press
ENTR
to proceed.
There are three levels of password protection, which correspond to operator,
maintenance and configuration functions (
4-6). Each level allows access to all of
the functions in the previous level.
IMPORTANT
There are more VARS available when using the password, 929, for configuration. Use
caution when pressing any buttons while in this setup. Any changes made may alter the
performance of the instrument or cause the instrument to not function properly. Note that
if there is an accidental change to a setup, press EXIT to discard the changes made.
Table 4-6: Password Levels
PASSWORD
LEVEL
MENU ACCESS ALLOWED
No password
Operator
All functions of the MAIN menu:
TEST, GEN,
initiate
SEQ , MSG, CLR
818
Maintenance
Access to Primary Setup and Secondary Setup Menus except for
VARS
and
DIAG
929
Configuration
Secondary SETUP Submenus
VARS
and
DIAG
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...