8.2. PREDICTIVE DIAGNOSTICS
The analyzer’s test functions can be used to predict failures by looking at trends in their
values (refer to Table 8-2) and by comparing them values recorded for them at the
factory and recorded on the
T100 Final Test and Validation Data Form
(Teledyne ML
P/N 04551) that was shipped with your analyzer.
A convenient way to record and track changes to these parameters is the internal data
acquisition system (DAS). Also, APICOM control software can be used to download
and record these data for review even from remote locations (refer to Section 5.3.2.8
describes APICOM).
Table 8-2: Predictive Uses for Test Functions
TEST FUNCTION
DAS
FUNCTION
CONDITION
BEHAVIOR
INTERPRETATION
EXPECTED
ACTUAL
PRES
SMPPRS
sample gas
Constant within
atmospheric
changes
Fluctuating
Developing leak in pneumatic system
Slowly
increasing
Flow path is clogging up.
- Check critical flow orifice & sintered filter.
- Replace particulate filter
Slowly
decreasing
Developing leak in pneumatic system to
vacuum (developing valve failure)
DRK PMT
DRKPMT
PMT output
when UV Lamp
shutter closed
Constant within
±20 of check-
out value
Significantly
increasing
PMT cooler failure
Shutter Failure
SO
2
Concentration
CONC1
At span with
IZS option
installed
Constant
response from
day to day
Decreasing
over time
Change in instrument response
Degradation of IZS permeation tube
Standard
configuration at
span
stable for
constant
concentration
Decreasing
over time
Drift of instrument response; UV Lamp
output is excessively low.
SAMP FL
SMPFLW
Standard
Operation
Stable
Slowly
Decreasing
Flow path is clogging up.
- Check critical flow orifice & sintered filter.
- Replace particulate filter
Fluctuating
Leak in gas flow path.
LAMP RATIO
LAMPR
Standard
Operation
Stable and near
100%
Fluctuating or
Slowly
increasing
UV detector wearing out
UV source Filter developing pin holes
Slowly
decreasing
UV detector wearing out
Opaque oxides building up on UV source
Filter
UV lamp aging
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...