SyriXus Series Pumps Installation and Operation Guide
Section 1 Introduction
1-6
EAR99 Technology Subject to Restrictions Contained on the Cover Page
Table 1-5 1000x Technical Specifications
CYLINDER CAPACITY
1015.0 mL
POWER REQUIREMENTS
a
(Mains voltage line cord is a
“Disconnect Device”)
100 ± 10 VAC, 1.5 A maximum
117 ± 12 VAC, 1.5 A maximum
234 ± 23 VAC, 0.75 A maximum
LINE FREQUENCY
50 or 60 Hz
LINE VOLTAGE NOISE TOLERANCE
1.7
nominal rms line voltage, 10 µsecond pulses, any phase
angle, random or repetitive
DIMENSIONS
PUMP
Width: 27.18 cm
Depth: 46.74 cm
Height: 103.60 cm
CONTROLLER
Width: 27.18 cm
Depth: 30.48 cm
Height: 13.59 cm
WEIGHT
b
PUMP
38.5 kg
CONTROLLER
2.96 kg
FLOW RATE RANGE
Refer to Figure 1-3.
FLOW RATE ACCURACY
c
± 0.5% (maximum 1.5 µL/min seal leakage)
FLOW RATE DISPLAY RESOLUTION
1.0 µL/min
ANALOG OUTPUT ACCURACY
d
± 1% of selected range
DISPLACEMENT RESOLUTION
25.38 nL /step
REFILL TIME
2.5 minutes
REFILL OR DEPRESSURIZATION RATE
1.0 µL/min to 408 mL/min at any pressure from 0 to 137.9 bar
PRESSURE RANGE
0.6895 to 137.9 bar
PRESSURE ACCURACY
± 0.5% of full scale at constant temperature
PRESSURE REPEATABILITY
e
± 0.5% of full scale within 48 hours at constant temperature
ZERO PRESSURE DRIFT
± 0.25% of full scale within 48 hours at constant temperature
PRESSURE DISPLAY RESOLUTION
6.895 kPa
AMBIENT TEMPERATURE RANGE
5 to 40 °C
TEMPERATURE DRIFT
± 0.12% of full scale/°C
HUMIDITY 95%
maximum
DEAD (HEADSPACE) VOLUME
f
11.0 ± 0.7 mL
POLLUTION DEGREE
2
INSTALLATION CATEGORY
II
MAXIMUM ALTITUDE
2000 m
a. Underwriters Laboratories (UL) has certified all SyriXus series syringe pumps with the exception of the 100 VAC versions.
b. A team lift is recommended when moving this instrument.
c. Using water at 137.9 bar and a temperature controlled environment at 30 °C.
d. The analog output is an optional accessory.
e. Pressure repeatability specification is based upon re-zeroing pressure transducer every 48 hours. Refer to Section 3.8.11
"ZERO PRESS" in Section 3 "Basic Programming and Operation" for re-zeroing procedure.
f.
Volume in and above the piston seal, head clearance at automatic shutoff, and inlet and outlet ports to the fittings.
Factory Set
}