Page 57
Testing the System with
SxS Pro
To test the RiverPro/RioPro using SxS Pro:
1. Start SxS Pro and establish communications with the RiverPro/RioPro ADCP.
For help on using
SxS Pro
, see the SxS Pro User’s Guide.
2. On the Tests menu, click ADCP Tests to verify the RiverPro/RioPro is functioning properly.
3. Click Run Tests.
4. At the end of the test, click the Stop PC2 button to end the PC2 test. Click Exit to exit the ADCP
Tests dialog.
The tests should be run while the RiverPro/RioPro ADCP is in water. Running the test in air
will not harm the RiverPro/RioPro, but some tests may fail in air.
Figure 20.
Testing the RiverPro/RioPro using SxS Pro
Testing the System with
BBTalk
To test the RiverPro/RioPro using BBTalk:
1. Interconnect and apply power to the system as described in
Setting up the RiverPro/RioPro Sys-
2. Start the BBTalk program.
3. Using BBTalk, send the RiverPro/RioPro the following commands: PS0, PS3, PA, and PC2.
For help on using