PS200
PORTABLE GAS MONITOR
CALIBRATION
64171, Rev. 11/December 2019
9-3
5. Once gas is applied, press and hold the R button to start the calibration. The 60 second count
-
down will commence (refer to
). After 60 seconds, the monitor
will be calibrated to the predefined values.
Figure 9-4: APPLY GAS Screen
6. The timer will count up to 9 seconds, during which time the monitor will adjust the gain to the
sensors accordingly. No manual input is required.
Figure 9-5: APPLY GAS Screen
7. Once calibration is successfully completed, the user has the option of saving the results (see
CAL COMPLETE/SAVE RESULTS Screen
).
Figure 9-6: CAL COMPLETE/SAVE RESULTS Screen
8. Using the L button, highlight one of the 3 the available options:
NOTE: The test gas must match the predefined cylinder values stored in the monitor.