8
64171LMP
Revision 3
PS200-LMP
PORTABLE GAS MONITOR
3.2.2.1. Filter Check
First, the Filter Check screen (see
) prompts the user to check the filters. If
filters have been checked, press
YES
to continue the monitor’s warm-up sequence. If
NO
is
selected, the monitor will automatically enter the 'switch off' sequence. Once the monitor shuts
down, check the filters and restart the monitor.
Figure 5: Filter Check
3.2.2.2. Flow Fault Check
Next, the FLOW FAULT TEST prompt screen appears (see
). Select
SKIP
to continue without performing a Flow Fault Test. If
OFF
is selected, the monitor will automatically
enter the 'switch off' sequence.
To perform a flow fault test, block probe tip with your finger for approximately 5 seconds. The
pump should flow fault, and the screen will turn red. Press
NEXT
to continue the monitor’s warm-
up sequence.
Figure 6: Flow Fault Test
3.2.3. Date and Time
The LCD displays the date and time during warm-up, as illustrated in
.
Flow Fault Test Prompt
Screen
Flow Fault Test
Successful