5. Select Readout Time from the Acquisition menu.
The readout time is reported on the Readout Time dialog as 2.08 ms. See
.
Figure 7-3: Typical WinX/32 Readout Time Dialog: 2.08 msec
6. Now go to the Hardware Setup
►
Controller/Camera tab.
•
Check the Custom Chip box.
•
Select the Custom Chip tab.
•
Make the following changes:
— Change Active to 20
— Change Pre-Dummies to 0
— Change Post-Dummies to 0
See
.
Figure 7-4: Typical WinX/32 Hardware Setup Dialog: Custom Chip Tab
7. Now go to the Hardware Setup
►
Controller/Camera tab.
8. Select Easy Bin from the Acquisition menu.
•
Select ALL. This will bin the 20 rows together.
9. Acquire a spectrum.
4
4
11-
01
26
_
007
3
4
4
11
-01
26
_007
4
Содержание ProEM Series
Страница 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Страница 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Страница 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 161: ...This page is intentionally blank ...