In LightField, the Custom Sensor pane, illustrated in
, is accessed by opening
the Sensor expander and clicking on the Custom Sensor button.
Figure 6-23: Typical LightField Custom Sensor Pane
By changing the values in the Active fields, you can increase image acquisition speed by
reducing the size of the active area in the definition. The result will be faster but lower
resolution data acquisition. Operating in this mode would ordinarily require that the
chip be masked so that only the reduced active area is exposed. This will prevent
unwanted charge from spilling into the active area or being transferred to the shift
register.
By default, if there are no Pre Dummy rows, the serial register will be cleared before
rows are shifted. If the Skip Serial Register Clean box is selected when there are no
Pre-Dummy rows, the register clean out will be skipped and the chip readout will be
faster. This feature is not available for PVCAM-supported cameras.
NOTE:
In LightField, the Clean Serial Register function only appears
in the Sensor Cleaning pane when the Inactive Area Top
Margin has been set to 0 rows. Deselect the check box to
deactivate the serial register cleaning.
4
4
11
-01
26
_
006
8
Содержание ProEM Series
Страница 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Страница 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Страница 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Страница 161: ...This page is intentionally blank ...