OLCT 60
FIXED POINTGAS MONITOR
USER MANUAL
NPO60GB
Revision F.2
51
Gas
Measuring
range
(ppm)
ADF
sensor
IS
sensor
Temperature
range
(°C)
% HR
Accuracy (ppm)
@20°C
Average
LifeExpe
ctancy
(months)
Resp. Time
T50/T90
(s)
Storage
time and
conditions
O3
Ozone
1.00
0 to +40 10 – 90
+/- 0.03 (from 0 to 0,2
ppm)
+/- 0.05 (from 0.2 to 1
ppm)
18
40/120
(c)
PH3
Phosphine
1.00
-20 to +40 20 – 90
+/- 0.05
18
30/120
(a)
SiH4
Silane
50.0
-20 to +40 20 – 95
+/- 1.0
18
25/120
(a)
SO2
Sulphur
Dioxide
10.0
30.0
100
-20 to +50 15 – 90
+/- 0.7 (range 0-10)
36
15/45
(a)
CMV
(
→
30/11/2018)
Vinyl chloride
200
0 to +40 15 – 90
+/- 5% (from 20 to 70%
FS)
24
10/50
(a)
CMV
(01/12/2019
→
)
200
-20 to +40 15 – 90
+/- 5% (from 20 to 70%
FS)
24
10/50
(a)
CH3Cl
Chloro-
methane
500
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
CH3Cl
Dichloro-
methane
500
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Freon R12
1 % vol.
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Freon R22
2000
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Freon R123
2000
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
FX56
2000
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Freon R134 a
2000
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Freon R134 a
2000
(IR)
-20 to +50 0 – 95
+/- 40 (from 0 to 50% FS)
+/- 100 (from 50 to 100%
FS)
60
40/170
(e)
Freon R11
1 % vol.
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
FS)
40
25/90
(d)
Freon R23
1 % vol.
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Freon R143 a
2000
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Freon R404 a
2000
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Freon R507
2000
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Freon R410 a
1000
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Freon R32
1000
-20 to +55 20 – 95
+/- 15% (from 20 to 70%
PE)
40
25/90
(d)
Содержание OLCT 60
Страница 1: ...NPO60GB Revision F 2 OLCT 60 FIXED POINTGAS MONITOR USER MANUAL ...
Страница 44: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL 38 NPO60GB Revision F 2 ...
Страница 50: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL 44 NPO60GB Revision F 2 ...
Страница 51: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL NPO60GB Revision F 2 45 8 Declaration of EU Conformity ...
Страница 52: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL 46 NPO60GB Revision F 2 ...
Страница 66: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL 60 NPO60GB Revision F 2 ...
Страница 67: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL NPO60GB Revision F 2 61 ...