089770000 Rev 05
Teledyne API N500 CAPS NO
X
Analyzer with NumaView™ Software
21
C
ONNECTING
C
OMMUNICATIONS
I
NTERFACES
E
THERNET
C
ONNECTION
For network or Internet communication with the analyzer, connect an Ethernet cable from
the analyzer’s rear panel Ethernet interface connector to an Ethernet port. Although the
analyzer is shipped with DHCP enabled by default, it should be manually configured with
a static IP address.
S
ERIAL
C
ONNECTION
Received from the factory, the analyzer COM port is set up for RS-232 communications
with data communication equipment (DCE). This port can be reconfigured for RS-232
communications with data terminal equipment (DTE) or with RS-485 by jumpering the
pins on JP1 as indicated in Table 2-5 (view/edit software settings Table 2-10).
WARNING – ELECTRICAL SHOCK HAZARD
Disconnect power before performing any operation that
requires entry into the interior of the analyzer. Contact
Technical Support (Section 5.9) before reconfiguring the
internal serial communications connector.
Figure 2-6. JP1 Location and Pin Arrangements
Table 2-5. JP1 Configurations for Serial Communication
Function
Jumpers
DSub Pins
2
3
DCE RS232
(default)
1-2, 4-5, 9-12
232Tx
232Rx
DTE RS232
2-3, 5-6, 9-12
232Rx
232Tx
RS485
7-8, 10-11, 9-12
485A(-) 485B(+)