Exposure Time > Readout Time
illustrates the timing diagram for this mode.
Figure 6-3: Timing Diagram: Rolling Shutter, Exposure Time > Readout Time
6.5
Fan Speed Control and Liquid Cooling
Fans are often used to remove heat from the camera due to their convenience.
However, under demanding conditions, fan vibration isolation is insufficient for single
molecule localization, high magnification imaging, or use with micro-manipulators.
Kuro solves this problem in two ways. First, a new, innovate fan mounting system has
been developed that isolates fan vibration from the rest of the camera. Side-by-side
testing with competing products indicate that Kuro outperforms competitors in terms
of vibration isolation.
To achieve the ultimate in low vibration performance, Kuro’s fan can be completely
turned off in software. Kuro provides liquid cooling ports for this purpose.
NOTE:
If the camera is inadvertently left on with the air and liquid
cooling disabled, a protection circuit will trip preventing the
camera from overheating. This circuit disables current to
the sensor cooling system until the camera returns. In this
case the camera is unable to cool the sensor. When setting
Kuro on a flat surface, be careful not to block the air vents.
4
4
11
-0
15
9
_
0016
Expose Out
Readout
Exposure Time
Configured in
LightField
Sensor
Exposure
Содержание Kuro
Страница 1: ...www princetoninstruments com Kuro sCMOS Camera System Manual 4411 0159 Issue 2 March 28 2019...
Страница 6: ...6 Kuro sCMOS Camera System Manual Issue 2 This page is intentionally blank...
Страница 18: ...18 Kuro sCMOS Camera System Manual Issue 2 This page is intentionally blank...
Страница 24: ...24 Kuro sCMOS Camera System Manual Issue 2 This page is intentionally blank...
Страница 34: ...34 Kuro sCMOS Camera System Manual Issue 2 This page is intentionally blank...
Страница 42: ...42 Kuro sCMOS Camera System Manual Issue 2 This page is intentionally blank...
Страница 47: ...This page is intentionally blank...