NDIR
Gas
Analyzer
Appendix
Teledyne Analytical Instruments., - Rev. 3
77
A-3.2 Codes for "s" Type Messages
The codes for "s" type messages are:
C
ODE
R
ETURN
M
ESSAGE
1C 1
st
(ir) Cal gas value
AP
Auto cal Period in hours
1R
chart Range 1 value
2R
chart Range 2 value
3R chart
Range 3 value
4R chart
Range 4 value
1H
Setpoint 2 trigger level
1L
Setpoint 1 trigger level
CD Cal Delay
1X
1
st
oXygen span cal setting
A-3.3 Codes for "c" Type Messages
The codes for "c" type messages are:
C
ODE
M
ODE
N
AME
V
ALUE
PL Panel Lock
1 = lock, 0 = unlock
DB
Display Brightness
0 thru 3 (3 is brightest)
RS
chart Range Selected
0 = Autorange, 1 thru 4
FS
Filter Selected
0 = filter 1
1 = filter 10
2 = filter 100 3 = filter 1000
CT Cal Track/hold
1 = track
0 = hold
TC Timed Cal type
0 = full
1 = Zero only
AE audible
Alarm Enable
0 = disabled 1 = enabled
LM Linear Mode
0 = linear 1 = nonlinear 2 = counts
CS Cal Select
0 = Zero 1 = Span 2 = both
AK
Alarm acKnowledge
1 = acknowledge alarm