Operator’s Manual
23
Performance Verification Procedure
This procedure can be used to verify the warranted characteristics of an HVD3000 or
HVD3000A probe. If the product does not meet specifications, it should be returned to a
Teledyne LeCroy service center. As there are no user accessible adjustments, there is no
adjustment procedure.
Required Equipment
The following table lists the test equipment and accessories (or their equivalents) that are
required for performance verification of the probe. This procedure is designed to minimize
the number of calibrated test instruments required. Because the input and output
connector types may vary on different brands and models of test instruments, additional
adapters or cables may be required. Only the parameters listed in boldface in the "Minimum
requirements" column must be calibrated to the accuracy indicated.
Description
Requirements
Example Equipment
Digital Multimeter
DC: 0.1% accuracy
AC: 0.2% accuracy to measure
7 mVrms to 37 Vrms at 70 Hz
Agilent 34401A
Fluke 8842A-09
Keithley 2001
Function Generator
Output sine wave:
20 Vpp HVD310x and HVD3602
37 Vrms HVD3605*
* May require an amplifier to obtain
required voltage, such as Tegam
2340.
Agilent 33120A
Stanford Research DS34
Leader LAG-120B
BNC Coaxial Cable
Male-to-Male 50 Ω Cable
Pomona 5697-36
Calibration Fixture
ProBus Extender Cable
Teledyne LeCroy PROBUS-CF01
Banana Plug Adapter
Female BNC-to-Dual
Banana Plug
Pomona 1269 Mueller BU-00260
Insulated Banana Plug
Insulated Banana Couplers (2)
Insulated BNC-to-Shrouded
Banana Plug
Mueller BU-5671-B-12-0
Mueller BU-32601-2 (R)
Mueller BU-32601-6 (B)
Содержание HVD3000
Страница 1: ...Operator s Manual HVD3000 HVD3000A High Voltage Differential Probes...
Страница 2: ......
Страница 3: ...HVD3000 HVD3000A High Voltage Differential Probes Operator s Manual November 2018...
Страница 16: ...HVD3000 HVD3000A High Voltage Differential Probes 10 HVD3106 6M HVD3106A 6M TYPICAL BANDWIDTH...
Страница 47: ......
Страница 48: ...930490 00 Rev A November 2018...