Performance Verification Procedure 

This procedure can be used to verify the warranted characteristics of an HVD3000 or 
HVD3000A probe. If the product does not meet specifications, it should be returned to a 
Teledyne LeCroy service center. As there are no user accessible adjustments, there is no 
adjustment procedure. 

Required Equipment 

The following table lists the test equipment and accessories (or their equivalents) that are 
required for performance verification of the probe. This procedure is designed to minimize 
the number of calibrated test instruments required. Because the input and output 

connector types may vary on different brands and models of test instruments, additional 
adapters or cables may be required. Only the parameters listed in boldface in the "Minimum 
requirements" column must be calibrated to the accuracy indicated. 

Description

 

Requirements

 

Example Equipment

 

Digital Multimeter 

DC: 0.1% accuracy 

AC: 0.2% accuracy to measure

  

7 mVrms to 37 Vrms at 70 Hz 

Agilent 34401A  

Fluke 8842A-09  

Keithley 2001 

Function Generator 

Output sine wave: 

20 Vpp HVD310x and HVD3602 

37 Vrms HVD3605* 
 

* May require an amplifier to obtain 

required voltage, such as Tegam 

2340. 

Agilent 33120A 

Stanford Research DS34  

Leader LAG-120B 

BNC Coaxial Cable 

Male-to-Male 50 Ω Cable 

Pomona 5697-36 

Calibration Fixture 

ProBus Extender Cable 

Teledyne LeCroy PROBUS-CF01 

Banana Plug Adapter 

Female BNC-to-Dual  

Banana Plug 

Pomona 1269 Mueller BU-00260 

Insulated Banana Plug 

Insulated Banana Couplers (2) 

Insulated BNC-to-Shrouded  

Banana Plug 

Mueller BU-5671-B-12-0  

Mueller BU-32601-2 (R)  

Mueller BU-32601-6 (B) 

Содержание HVD3000

Страница 1: ...Operator s Manual HVD3000 HVD3000A High Voltage Differential Probes...

Страница 2: ......

Страница 3: ...HVD3000 HVD3000A High Voltage Differential Probes Operator s Manual November 2018...

Страница 4: ...ctly prohibited However clients are encouraged to duplicate and distribute Teledyne LeCroy documentation for their own internal educational purposes Teledyne LeCroy is a trademark of Teledyne LeCroy I...

Страница 5: ...quired Equipment 23 Preliminary Procedure 24 Certification Procedure 24 Checking Probe Attenuation in XStreamBrowser 26 HVD3000 HVD3000A Test Record 27 Operation 28 Connecting to the Test Instrument 2...

Страница 6: ...y product returned to one of our authorized service centers within this period However in order to do this we must first examine the product and find that it is defective due to workmanship or materia...

Страница 7: ...mall control signals floating on high common mode voltages The maximum rated differential voltage specifications are provided for using each probe within a wide differential voltage range The probes w...

Страница 8: ...r the widest range of V div settings Compatibility HVD3000 and HVD3000A probes are compatible with most Teledyne LeCroy MAUI oscilloscopes equipped with the ProBus interface See the HVD3000 product pa...

Страница 9: ...ent if anti static measures are not taken Precautions WARNING To avoid personal injury or damage due to electric shock orfire Do not overload observe all terminal ratings Do not apply any potential th...

Страница 10: ...ational environment listed Do not use in wet or explosive atmospheres Keep product surfaces clean and dry Do not operate with suspected failures Before each use inspect the probe and accessories for a...

Страница 11: ...tegory per IEC EN 61010 031 2015 CAUTION The operating altitude of the probe is derated to 2000 m 6560 ft when used with the above accessories WARNING Each accessory has a different measurement overvo...

Страница 12: ...ews the overall length is 63 mm 2 48 inches 4 mm Banana female connector 1 ea red black Insulated 1000 V CAT III PK HVA 05 2 Safety Alligator Clips 2 Designed to reliably grip large components such as...

Страница 13: ...ed with a flexible stem to access deep targets in dense environments and a 4 5 mm 0 177 inch hook to attach to wire leaded parts The overall length is 157 6 mm 6 20 inches 4 mm Banana female connector...

Страница 14: ...10 90 14 ns 2 9 ns 4 4 ns CMRR Test Limits 23 C 80 db 50 Hz 60 db 1 MHz 80 db 50 Hz 60 db 1 MHz 80 db 50 Hz 60 db 1 MHz ELECTRICAL CHARACTERISTICS ALL PROBES Max Rated Diff Voltage between each input...

Страница 15: ...Operator s Manual 9 HVD310x HVD310xA Bandwidth HVD3102 HVD3102A TYPICAL BANDWIDTH HVD3106 HVD3106A TYPICAL BANDWIDTH...

Страница 16: ...HVD3000 HVD3000A High Voltage Differential Probes 10 HVD3106 6M HVD3106A 6M TYPICAL BANDWIDTH...

Страница 17: ...Operator s Manual 11 HVD310x HVD310xA Common Mode Rejection Ratio HVD3102 HVD3102A TYPICAL CMRR PERFORMANCE HVD3106 HVD3102A TYPICAL CMRR PERFORMANCE...

Страница 18: ...andling the inputs Handling the inputs while connected to active signals above this burn limit could result in injury to the operator WARNING To avoid risk of electric shock or fire comply with the bu...

Страница 19: ...m 5 1 inches the jaw opens to 20 mm 0 79 inch maximum 4 mm Banana female connector 1 ea red black 2000 V DC Peak AC CAT I 1000 Vrms CAT III 1500 Vdc CAT III PK HVA 07 2 Case Soft storage case SAC 01A...

Страница 20: ...b 1 MHz ELECTRICAL CHARACTERISTICS ALL PROBES Max Rated Diff Voltage between each input 2000 V DC Peak AC Max Measurable Diff Voltage before saturation 2000 V DC Peak AC Max Common Mode Voltage either...

Страница 21: ...Operator s Manual 15 HVD3206 HVD3206A Bandwidth HVD3206 HVD3206A TYPICAL BANDWIDTH HVD3206 6M HVD3206A 6M TYPICAL BANDWIDTH...

Страница 22: ...HVD3000 HVD3000A High Voltage Differential Probes 16 HVD3206 HVD3206A Common Mode Rejection Ratio TYPICAL CMRR PERFORMANCE...

Страница 23: ...ts Handling the inputs while connected to active signals above this burn limit could result in injury to the operator WARNING To avoid risk of electric shock or fire comply with the burn limit when me...

Страница 24: ...e overall length is 116 mm 4 57 inches and the jaw opens to 22 mm 0 87 inch 1 ea red black 6000 V CAT I 1000 Vrms CAT III 1500 Vdc CAT III PK HVA 06 2 Case Soft storage case SAC 01A 1 Foam Custom foam...

Страница 25: ...S ALL PROBES Max Rated Diff Voltage between each input 7000 V DC Peak AC Max Measurable Diff Voltage before saturation 7600 V DC Peak AC Max Common Mode Voltage either input to ground 7600 V DC Peak A...

Страница 26: ...HVD3000 HVD3000A High Voltage Differential Probes 20 HVD3605 HVD3605A Bandwidth TYPICAL BANDWIDTH HVD3605 HVD3605A Common Mode Rejection Ratio TYPICAL CMRR PERFORMANCE...

Страница 27: ...ts Handling the inputs while connected to active signals above this burn limit could result in injury to the operator WARNING To avoid risk of electric shock or fire comply with the burn limit when me...

Страница 28: ...of the CAL out signal For oscilloscopes with the CAL signal on a BNC connector a BNC to Banana adapter e g Pomona model 1296 may beused 5 Press AUTOSETUP 6 Set the sensitivity of the probe to1 V Div 7...

Страница 29: ...nal adapters or cables may be required Only the parameters listed in boldface in the Minimum requirements column must be calibrated to the accuracy indicated Description Requirements Example Equipment...

Страница 30: ...n generator to sine wave 70 Hz and an output voltage of approximately 7 Vrms into a high impedance output for HVD310x HVD310xA HVD3206 and HVD3206A 37 Vrms for HVD3605 and HVD3605A 2 Set the DMM to me...

Страница 31: ...ificant places x xx on line 13 in the Test Record 14 Verify that the error 1 00 15 Decrease the oscilloscope scale factor to 5 V Div 16 Record the DMM reading to 0 01 mV resolution in the Test Record...

Страница 32: ...icon 2 From the browser menu bar choose File Connect to Local Instrument 3 Navigate to LeCroy XStreamDSO Acquisition C1 InputB HVDxxxxx where xxxxx is your probe model number The value is shown next...

Страница 33: ...ith a DMM or oscilloscope Thus the generator is not required to be calibrated Step Description Intermediate Data Test Result 5 Function Generator Output Voltage V 11 Probe Output Voltage V 13 Gain Err...

Страница 34: ...rmance capability exercise care when connecting the probe Increasing the parasitic capacitance or inductance in the input paths may introduce a ring or slow the rise time of fast signals Input leads t...

Страница 35: ...ircuits being measured as they cannot provide adequate protection if they come in contact with electrical sources Operating with an Oscilloscope When the probe is connected to a Teledyne LeCroy oscill...

Страница 36: ...set for the probe and oscilloscope combination can then be calculated as Max Positive Offset Available Max Positive Offset Probe and Oscilloscope together Oscilloscope Positive Offset at Oscilloscope...

Страница 37: ...OTE Because AC coupling is on the probe output DC voltages beyond the probe s linear range will cause the amplifier to saturate and inaccurately render the displayed waveform Do not exceed the maximum...

Страница 38: ...to the output If the probe is set to 50x attenuation and the ambient temperature changes by 10 C the DC offset drift could be as high as 70 V C 50 10 C 35 mV referred to the probe tip If measuring a 3...

Страница 39: ...st be returned to a Teledyne LeCroy service facility for diagnosis and exchange CAUTION Do not remove the covers Refer all servicing to qualified personnel A defective probe under warranty will be rep...

Страница 40: ...manual 3 Label the case with a tag containing The RMA Name and address of the owner Product model and serial number Description of failure or requisite service 4 Pack the product case in a cardboard s...

Страница 41: ...white papers and videos are available to help you get the most out of your Teledyne LeCroy products The Datasheet published on the product page contains the detailed product specifications Service Ce...

Страница 42: ...ning the CMRR is the relative gain matching between the and input paths Obtain high CMRR values by precisely matching the input attenuators in a differential amplifier The matching includes the DC att...

Страница 43: ...l Mode Range and Common Mode Range Differential Mode range is the maximum signal that can be applied between the and inputs without overloading the amplifier which otherwise would result in clipping o...

Страница 44: ...plies to test and measuring circuits connected to the distribution part of the building s low voltage mains installation Measurement Category II CAT II applies to test and measuring circuits connected...

Страница 45: ...use EM interference 2 Emissions exceeding the levels required by this standard may occur when product is connected to a test object 3 To ensure compliance with the applicable EMC standards use high qu...

Страница 46: ...easurement and Test Canadian Certification The probe has been certified by Underwriters Laboratories UL to conform to the following safety standard and bears cUL Listing Mark CAN CSA C22 2 No 61010 03...

Страница 47: ......

Страница 48: ...930490 00 Rev A November 2018...

Отзывы: