126B - PAGE 16
SYMPTOM:
SYMPTOM:
SYMPTOM:
SYMPTOM:
SYMPTOM: Sensor has proper resistance readings, but little or no output with flow.
CAUSE:
CAUSE:
CAUSE:
CAUSE:
CAUSE: Plugged sensor.
A
A
A
A
ACTION:
CTION:
CTION:
CTION:
CTION: Shut off gas supply and power supply. Remove cover and PC board from unit. Remove
sensor assembly and examine. If sensor has evidence of plugging, clean or replace as applicable.
SYMPT
SYMPT
SYMPT
SYMPT
SYMPTOM:
OM:
OM:
OM:
OM: Flowmeter reads other than 0.00 VDC with no flow, or there is a small flow when
flowmeter reads 0.00 VDC.
CA
CA
CA
CA
CAUSE:
USE:
USE:
USE:
USE: ZERO potentiometer is out of adjustment.
A
A
A
A
ACTION:
CTION:
CTION:
CTION:
CTION: Shut off all flow. Adjust ZERO potentiometer until output reads 0.00 VDC.
SYMPTOM:
SYMPTOM:
SYMPTOM:
SYMPTOM:
SYMPTOM: Flowmeter out of calibration and non-linear.
CAUSE:
CAUSE:
CAUSE:
CAUSE:
CAUSE: Leaks in gas inlet or outlet fittings.
A
A
A
A
ACTION:
CTION:
CTION:
CTION:
CTION: Check all fittings for leaks by placing soap solution on all fittings between gas supply and
final destination of gas. Check flowmeter for leaks. Replace “O” rings if required or recalibrate as
necessary.
Содержание HASTINGS 200 Series
Страница 12: ...126B PAGE 12 ...
Страница 18: ...126B PAGE 18 ...
Страница 20: ...126B PAGE 20 ...
Страница 21: ...126B PAGE 21 7 0 Diagrams and Drawings ...
Страница 22: ...126B PAGE 22 PC 810Assembly InformationfromHastingsdrawing30837rev F ...
Страница 23: ...126B PAGE 23 PC 810Schematic InformationfromHastingsdrawing30837rev F ...
Страница 24: ...126B PAGE 24 ...
Страница 25: ...126B PAGE 25 ...
Страница 26: ...126B PAGE 26 ...
Страница 27: ...126B PAGE 27 ...
Страница 28: ...126B PAGE 28 ...