There are 4 relay alarm outputs (AL1-AL4) on the rear panel. AL1 is for system
okay, AL2 is for concentration limit 1 exceeded and AL3 is for concentration limit 2
exceeded. The relay alarm output AL1 is enabled all the time, whereas the AL2 and
AL3 can be enabled/disabled by going into the diagnostics menu (with 929
password), then factory options, then turning conc. Alarm Relays ON or OFF. When
the concentration alarm relays are enabled, concentration alarms status is no longer
available through the “Status output” pins. So, either the alarm relays or the status bits
could be used to monitor the concentration alarms but not both at the same time. The
default factory setting is to enable the concentration alarm relays. The AL1 relay is
energized when the system is okay and de-energized when the system has a fault. The
AL2 and AL3 relays energize when the corresponding concentration limits are
exceeded.

11.1.1.  Interpreting WARNING Messages

The most common and/or serious instrument failures will result in a warning message
being displayed on the front panel. The table below lists warning messages, along with
their meaning and recommended corrective action.
It should be noted that if more than two or three warning messages occur at the same
time, it is often an indication that some fundamental analyzer sub-system (power supply,
Relay Board, Motherboard) has failed rather than indication of the of the specific failures
referenced by the warnings. In this case, it is recommended that proper operation of
power supplies, the Relay Board, and the A/D Board be confirmed before addressing the
specific warning messages.
The analyzer will alert the user that a Warning Message is active by displaying the
keypad label 

MSG 

on the Front Panel.  In this case the Front panel display will look

something like the following:

The analyzer will also alert the user via the Serial I/O COM port(s) and cause the FAULT
LED on the Front panel to blink.
To View or Clear the various warning messages press:

SAMPLE       RANGE=50.00 PPM                CO = 00.00
<TST  TST>    CAL                    

 MSG

       CLR  SETUP

Содержание GFC7000EU

Страница 1: ...584 05134 Rev 1 MODEL GFC7000EU C O2 ANALYZER Addendum to GFC7000E Manual PN 04584 Teledyne Analytical Instruments 16830 Chestnut St City of Industry Ca 91748 TEL 626 961 9221 FAX 626 961 2538 TEL 626 934 1500 E MAIL ai customerservice teledyne com WEB SITE http www teledyne ai com ...

Страница 2: ...rence manual covering the details of the instrument s components Only the sections of the manual that are different between the GFC7000E and a GFC7000EU are listed in this addendum 2 SPECIFICATIONS AGENCY APPROVALS AND WARRANTY 2 1 Specifications Table Model GFC7000EU Specifications Ranges User selectable to any full scale range from 0 100 ppb to 0 100 ppm Measurement Units ppb ppm µg m3 mg m3 use...

Страница 3: ...us Outputs 8 Status outputs from opto isolators Control Inputs 6 Control Inputs 2 defined 4 spare I O One 1 RS 232 One 1 RS 485 RS 232 Ethernet Baud Rate 300 115200 Certifications CE EN61010 1 90 A1 92 A2 95 EN61326 Class A 1 As defined by the USEPA 2 At constant temperature and voltage 3 Or 0 2 ppm whichever is greater 4 Or 0 1 ppm whichever is greater 5 Above 10 ppm range otherwise 0 2 ppm for l...

Страница 4: ...centration limits are exceeded 11 1 1 Interpreting WARNING Messages The most common and or serious instrument failures will result in a warning message being displayed on the front panel The table below lists warning messages along with their meaning and recommended corrective action It should be noted that if more than two or three warning messages occur at the same time it is often an indication...

Страница 5: ...zer s CPU or Mother Board can result in any or ALL of the following messages Table Warning Messages Warning Message Fault Condition Possible Causes BENCH TEMP WARNING The optical bench temp is controlled at 48 5 C Bench Heater Bench Temperature Sensor Relay controlling the Bench Heater Entire Relay Board I2 C Bus BOX TEMP WARNING Box Temp is 5 C or 48 C Box Temperature typically runs 7o C warmer t...

Страница 6: ...RESS_SET variable Loss of purge flow Failed purge pressure sensor Purge pressure regulator not set properly SAMPLE PRES WARN Sample Pressure is 15 in Hg or 35 in Hg Normally 29 92 in Hg at sea level decreasing at 1 in Hg per 1000 ft of altitude with no flow pump disconnected If Sample Pressure is 15 in HG Blocked Particulate Filter Blocked Sample Inlet Gas Line Failed Pressure Sensor circuitry If ...

Страница 7: ...er Fatal Error caused software to restart Loose connector wiring WHEEL TEMP WARNING The filter wheel temperature is controlled at 68 5 C Blocked Cooling Vents below GFC Assembly Analyzer s Top Cover removed Wheel Heater Wheel Temperature Sensor Relay controlling the Wheel Heater Entire Relay Board I2 C Bus APPENDIX B GFC7000EU Spare Parts List NOTE Use of replacement parts other than those supplie...

Страница 8: ...M300E B17 033560000 Field Mirror 32 Pass M300E 035950100 Disk On Chip with Software 036110400 Sync Demodulator Board 036310000 4 20 mA Output PCA 037250000 Assembly Heater Optical Bench 039750000 Keyboard 040010000 ASSY Fan Rear Panel E Series 040030300 PCA Press Sensors w SW49 035280000 CO2 Scrubber Assembly 040690200 Mother Board relay alarm outputs 040880000 Wheel Position Sensor Assembly 04135...

Страница 9: ...V 039060000 Assy Pressure Transducer M360U GFC7000 Table B 2 GFC7000E Expendables List Part Number Description Qty 009600400 M300E Expandables Kit Includes This kit contains the following items 009690100 Akit TFE Fltr Elements 47mm 5um 25 1 FL0000001 Sintered Filter 002 024900 1 HW0000020 Spring Flow Control 1 OR0000001 O Ring Flow Control 2 PU0000022 Pump Rebuild Kit KNF Model NO5ATI 1 Note01 23 ...

Отзывы: