Teledyne Foxy 200
4 Using the control module
above the negative of the
Slope Stop
value. If only one of the stop
signal conditions is fulfilled the fraction collection will not stop.
Start L+S End L or S
- The fraction collection:
l
starts when both start signal conditions are fulfilled simultaneously,
detector signal increases above
Level Start
value and also the first
derivative of the detector signal increases above
Slope Start
value. If
only one of the start signal conditions is fulfilled the fraction collection
will not start.
l
stops when any of the stop signal conditions is fulfilled (whatever
occurs first), detector signal decreases below
Level Stop
value or the
first derivative of the detector signal first decreases and then increases
above the negative of the
Slope Stop
value. If none of the stop signal
conditions is fulfilled the fraction collection will not stop before
Stop
Time
is reached.
Start L or S End L+S
- The fraction collection:
l
starts when any of the start signal conditions is fulfilled (whatever
occurs first), detector signal increases above
Level Start
value or the
first derivative of the detector signal increases above
Slope Start
value. If none of the start signal conditions is fulfilled the fraction
collection does not start.
l
stops when both stop signal conditions are fulfilled simultaneously,
detector signal decreases below
Level Stop
value and also the first
derivative of the detector signal first decreases and then increases
above the negative of the
Slope Stop
value. If only one of the stop
signal conditions is fulfilled the fraction collection does not stop.
Start L or S End L or S
- The fraction collection:
l
starts when any of the start signal conditions is fulfilled (whatever
occurs first), detector signal increases above
Level Start
value or the
first derivative of the detector signal increases above
Slope Start
value. If none of the start signal conditions is fulfilled the fraction
collection does not start.
l
stops when any of the stop signal conditions is fulfilled (whatever
occurs first), detector signal decreases below
Level Stop
value or the
first derivative of the detector signal first decreases and then increases
above the negative of the
Slope Stop
value. If none of the stop signal
conditions is fulfilled the fraction collection does not stop before
Stop
Time
is reached.
Filter
Sets the data filter on the detector signal defined on the given row of
the fraction table. Default value of
1
means no filtering of data will be
- 10 -