User Manual
CP Series
Version 3.1 – June 2020
53
6.3.16 E149:
C
AR
.
POINTER
The carousel is in the laser pointer position. Making an exposure with the carousel in the laser pointer
position may damage the electronics of the laser pointer. The laser pointer would also show up in the
developed film. Put the carousel in a correct position to continue with the exposure.
6.3.17 E150:
C
AR
.
U
NLOCKED
The carousel is in between 2 positions. Verify that ‘click’ can be heard and that the carousel is in a good
position.
6.3.18 E
151:
C
AR
.
B
E
50
K
V
The carousel is in the ‘be window’ position but a kV is selected of more than 50kV. Select a kV below or
equal to 50kV or use the ‘full open with 3mm aluminum filter’ instead.