
b) Measuring Procedure
Temperature Measurem.: K-type thermocouple (
) for waste or flue gas temperature
for air or ambient temperature.
O -Measurement :
Electrochemical measuring cell.
CO-Measurement :
Electrochemical measuring cell.
Draft Measurement :
Piezo-resistive principle with internal temperature compensation.
Measuring Duration:
Short-term memory measurements of max. 60 minutes are possible, followed by
a new calibration phase with ambient air.
Waste Gas Measurement : Via an external water separator and filter, the waste gas is fed to the sensors by
means of a gas feed pump. The pump capacity during the feeding phase is
approx. 0.8 l/min.
Sensor Calibration:
60 seconds after switching on the instrument.
CO Concentration:
CO sensor with H compensation, measuring range 0 - 4.000 ppm. Cutoff
threshold at 4.000 ppm for sensor protection via separate flush pump.
The remaining measuring values are not affected. The instrument is switched on
again at a value of 1.600 ppm.
Waste Gas Sampling:
By means of a waste gas sampling probe with retainer cone.
NiCr-Ni
K-type thermocouple (NiCr-Ni)
2
2
3
Содержание CEA 9001
Страница 1: ... US Version V3 1 C E A ombustion fficiency nalyzer US Version V3 1 ...
Страница 3: ......
Страница 13: ...10 4 Operating Elements 0 I START STOP S1 S2 S3 ENTER mbar hPa ENTER ...
Страница 14: ...11 ...
Страница 17: ...14 Natural Gas ...
Страница 18: ...15 ...
Страница 23: ...20 Cursor flashes OK to key in your entry ...
Страница 24: ...Intermediate storage of the graphics and allmeasured values 21 ...
Страница 26: ...23 ABC abc0123 ...
Страница 28: ...25 No Files are existend ...
Страница 30: ...27 ...
Страница 31: ...28 ...
Страница 35: ...32 ...
Страница 36: ...33 Attention All memories inlcuding additional data will be deleted ...
Страница 38: ...35 Adjustable display illumination from 0 90 seconds in increments of 5 seconds ...
Страница 39: ...36 Attention Confirming Reset with ENTER cancels all instrument settings All memory blocks are deleted ...
Страница 45: ...42 ...
Страница 46: ...43 ...