ADFM® Flow Meter
Section 3 Maintenance and Troubleshooting
3-6
3.4 Troubleshooting the
ADFM
The following sections provide assistance in locating and iso-
lating faults in the system.
3.4.1 Troubleshooting
Tables
Troubleshooting information for system power-up is provided in
Table 3-1, and for system operation in Table 3-2. Possible
problems and actions are listed in order of likelihood.
Note
Call the factory for assistance before replacing the mother-
board.
Table 3-1 Troubleshooting on Power-Up
Symptom
Possible Problem(s)
Action(s)
No wakeup message on LCD or com-
puter screen
Power
Motherboard
Verify battery voltage and connec-
tions. Verify connection to external
DC supply.
Contact the factory.
No wakeup message on LCD, but com-
puter displays wakeup message when a
manual reset is performed.
LCD
Front panel interface board
Cable connection between LCD and
front panel
Cable connection between front
panel interface board and mother-
board.
Motherboard
Replace LCD
Replace front panel interface board
Replace ADFM Motherboard.
No wakeup message on computer
screen, but LCD displays wakeup mes-
sage and completes Built-In-Tests (BIT).
RS-232 connection
Incorrect port settings on computer
Front panel interface board
Motherboard
Troubleshoot cabling
Adjust communication port settings
on computer
Replace front panel interface board
Replace motherboard.
Содержание ADFM Hot Tap
Страница 2: ......
Страница 4: ......
Страница 20: ...ADFM Flow Meter Section 1 Introduction 1 10 ...
Страница 85: ...ADFM Flow Meter Section 2 Setup and Installation 2 65 Figure 2 73 Viewing data in Expert View ...
Страница 86: ...ADFM Flow Meter Section 2 Setup and Installation 2 66 ...
Страница 98: ...ADFM Flow Meter Section 3 Maintenance and Troubleshooting 3 12 ...
Страница 114: ......
Страница 116: ......