1. Linear oxygen (0.1 to 100.0%)
*
2. Low range linear oxygen (10 to 10,000 ppm)
The output spans are adjustable in Set-up steps 21 and 22.
21. TRANSMITTER ZERO CHANNEL 1
Select transmitter zero for output Channel 1.
Range 0.0 to 99.9% Oxygen.
Range 0 to 0.999% (9,990ppm) Oxygen.
Default setting is 0.0%.
22. TRANSMITTER SPAN CHANNEL
Select transmitter span for output Channel 1.
Linear Oxygen-
Range 0.1 to 100.0% Oxygen. Default setting is 10.0%.
Low Oxygen-
Range 0.001 to 1.000% (10 to 10,000ppm) Oxygen. Default setting is 0.100%.
23. TRANSMITTER OUTPUT CHANNEL 2
Select the type of output required for Channel 2.
Options:
1. Sample Gas, oxygen
2. Linear (Low) Oxygen, 0.1 to 100.0 % *
3. Logarithmic oxygen, 0.1 to 20 %
4. % carbon dioxide
5. Reducing oxygen %
6. Oxygen sensor EMF
24. TRANSMITTER ZERO CHANNEL 2
The output zero and span of Channel 2 is set in set-up steps 24 and 25. Range limits are shown below.
25. TRANSMITTER SPAN CHANNEL 2
Output
Zero Range
Span Range
Minium span
SAMPLE GAS, OXYGEN
0.0 to 99.9 %
0.1 to 100 %
0.1 %
LOW OXYGEN
0.0 to 99.9 %
0.1 to 100 %
0.1 %
LOG OXYGEN
0.1 % oxygen fixed
20 % oxygen fixed
(see Note 1)
CARBON DIOXIDE
0 to 80 %
20 to 100.0 %
20 %
REDUCING OXYGEN
10
-
1
to 10
-28
%
10
-
2
to 10
-
30
% Two
decades
(see Note 2)
oxygen in one
oxygen in one
decade steps
decade steps.
SENSOR EMF
0 to 1100 mV in
100 to 1300 mV
100 mV
100 mV steps
in 100 mV steps
NOTE
1:
For log oxygen scale details, Refer to Appendix 2.
2:
Note that the reducing oxygen span is shown on the display as the exponent only. -1 represents 10
-1
% oxygen.
26. SAMPLE MODE
Options
1. Fast Sample
2. Display Sample
*
3. Continuous
Page
34
Teledyne Analytical Instruments 9070 Oxygen Analyzer