Section 2: Sensor and Sampling Techniques
page 7
2.3.2 Extractive Installation
For extractive installations we recommend our sample sys-
tem ESS, which may be equipped with a variety of features,
such as: isolation valve, coalescing or particulate filter,
pressure regulator, calibration sample injection or extraction
port, pressure gauge, flow meter, weatherproof enclosure.
Refer to the ESS literature for more information.
If the resources to make your own sample system are avail-
able, the following two diagrams may be used as a guideline
to configure a simple system.
8800T
4-20mA loop cable
Exhaust
Regulator or
Needle V
alv
e
Safety shut-off Valve
Extractive Installation, Sensor
Measuring at Line Pressure
8800T
4-20mA loop cable
Exhaust
Regulator or
Needle Valve
Extractive Installation, Sensor
Measuring at Ambient Pressure
Sample Cell
oint Transmitter
Ma
Se
Mo
Think
Safety
Read
Man
ual
www.xentaur.com
Sample Cell
oint Transmitter
Ma
Se
Mo
Think
Safety
Read
Man
ual
www.xentaur.com
Main
Gas Line
Main
Gas Line