![background image](http://html.mh-extra.com/html/teledyne/6712sr/6712sr_installation-and-operation-manual_1080774087.webp)
6712SR Refrigerated Sampler
Section 4 Standard Programming
4-33
Table 4-5 Programming Example With 750 Module Installed
1
MODULE INSERTED---
DOWNLOAD DATA NOW
OR LOSE ALL DATA!
DONE
With the sampler off, insert the module in the bay on the controller.
Turn the sampler on by pressing the
On/Off
key. Press
↵
. Download stored
data, if you want to keep it. Pressing DONE erases it and advances to the
next screen.
2
6712 SAMPLER
STANDARD PROGRAMMING
For HELP at any
screen press ? key.
This screen will now appear for 8 seconds.
3
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
The option PROGRAM will be blinking. Press
↵
.
4
SITE DESCRIPTION:
“FACTORY051“
CHANGE?
YES
NO
The option NO will be blinking. Press
↵
.
For the purposes of this example, it isn’t necessary to change the description.
5
SELECT UNITS FOR
FLOW RATE:
cfs
gps gpm Mgd
lps m3s m3h m3d
Select the desired units for the flow rate. For this example, choose cfs.
Press
↵
.
6
SELECT UNITS FOR
FLOW VOLUME:
cf gal
Mgal
m3 lit
Select the desired units for the flow volume. For this example, choose Mgal.
Press
↵
.
7
SELECT UNITS FOR
VELOCITY:
fps
mps
Select the desired units for the flow velocity. For this example, choose fps.
Press
↵
.
8
PROGRAM MODULE?
YES
NO
Select YES to program the module.
9
MODE OF OPERATION:
FLOWMETER
LEVEL AND VELOCITY
The 750 can have two different modes of operation: Flowmeter or Level and
Velocity.
For this example, choose Flowmeter. Press
↵
.
10
FLOW CALCULATION:
AREA * VELOCITY
LEVEL TO FLOW RATE
When you select Flowmeter, you will have two options for performing the
flow conversion: Area * Velocity or Level to Flow Rate.
For this example, choose Area * Velocity. Press
↵
.
11
LEVEL TO AREA:
CHANNEL SHAPE
DATA POINTS
For this example, choose Channel Shape. Press
↵
.
12
ROUND PIPE
U-CHANNEL
RECTANGULAR CHANNEL
TRAPEZOIDAL CHANNEL
For the channel shape, select Round Pipe. Press
↵
.
Содержание 6712SR
Страница 2: ......
Страница 4: ......
Страница 14: ...6712SR Refrigerated Sampler Table of Contents xii ...
Страница 28: ...6712SR Refrigerated Sampler Section 1 Introduction 1 14 ...
Страница 42: ...6712SR Refrigerated Sampler Section 2 Installation Preparation 2 14 ...
Страница 54: ...6712SR Refrigerated Sampler Section 3 Getting Started 3 12 ...
Страница 140: ...6712SR Refrigerated Sampler Section 6 SDI 12 Sondes 6 8 ...
Страница 188: ...6712SR Refrigerated Sampler Appendix A Menu Flowcharts A 24 ...
Страница 196: ...6712SR Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Страница 206: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 2 ...
Страница 207: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 3 ...
Страница 208: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 4 Locknut 10 32 Screw 10 32 x 3 4 ...
Страница 209: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 5 ...
Страница 210: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Страница 211: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 7 ...
Страница 212: ...6712SR Refrigerated Sampler Appendix D Replacement Parts D 8 ...
Страница 218: ...6712SR Refrigerated Sampler Appendix E Accessories List E 6 ...
Страница 222: ...6712SR Refrigerated Sampler Index Index 4 ...
Страница 224: ......