6712FR Refrigerated Sampler
Section 7 Remote Operation
7-10
Disable
This command is similar in functionality to a pin F low signal (a signal typically
used to enable or disable the sampler from a flow meter). At run time, the sam-
pler starts with the phone enable in the enabled state. PHONE DISABLE is
valid while a program is running. If successful, a reply of REMOTE ENABLE IS
FALSE will be returned.
A Remote Disable is the highest priority disable. If the sample is disabled with
this command, no condition other than command 6#, ENABLE, will allow the
program to become enabled again.
Enable
This command is used to re-enable a phone-disabled sampler. PHONE
ENABLE is valid while a program is running. If successful, a reply of REMOTE
ENABLE IS TRUE will be returned.
Grab Sample
This command causes the sampler currently running a program to take a sam-
ple. The sample will count as one of the program’s samples and is placed in
the current bottle (Part A’s current bottle when using two-part programming).
TAKE_SAMPLE is valid while the program is running, after the start time, but
not while in the MANUAL PAUSE screens. If successful, a response similar to
PUMPING XXML SAMPLE INTO BOTTLE ## will be returned. If not, a
response similar to NO SAMPLE, PROGRAM WAITING START will be
returned.
Table 7-4 6712 Remote Phone Commands (Continued)
Phone
Command
Name
Description
5
#
6
#
7
#
Содержание 6712FR and
Страница 2: ......
Страница 4: ......
Страница 8: ...6712FR Refrigerated Sampler Safety vi ...
Страница 16: ...6712FR Refrigerated Sampler Table of Contents viii ...
Страница 56: ...6712FR Refrigerated Sampler Section 3 Programming Introduction 3 12 ...
Страница 142: ...6712FR Refrigerated Sampler Section 6 SDI 12 Sondes 6 8 ...
Страница 174: ...6712FR Refrigerated Sampler Section 8 Maintenance 8 22 ...
Страница 198: ...6712FR Refrigerated Sampler Appendix A Menu Flowcharts A 24 ...
Страница 206: ...6712FR Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Страница 216: ...6712FR Refrigerated Sampler Appendix C General Safety Procedures C 10 ...
Страница 218: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 2 ...
Страница 219: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 3 ...
Страница 220: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 4 ...
Страница 221: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 5 ...
Страница 222: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 6 Locknut 10 32 Screw 10 32 x 3 4 ...
Страница 223: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 7 ...
Страница 224: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 8 ...
Страница 225: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 9 ...
Страница 226: ...6712FR Refrigerated Sampler Appendix D Replacement Parts List D 10 ...
Страница 232: ...6712FR Refrigerated Sampler Appendix E Accessories List E 6 ...
Страница 236: ...6712FR Refrigerated Sampler Index Index 4 ...
Страница 238: ......
Страница 240: ......
Страница 242: ...Warranty ...
Страница 244: ......