6712 Sampler
2-8
Note
The suction lines float when filled with air from the purge
cycles and when installed at depths exceeding the
maximum depths listed here. Secure the lines when
installing them at depths greater than the maximums.
Teflon suction line is compatible only with the 1 inch
diameter, stainless steel strainer.
2.5 Connecting An External
Instrument
2.5.1 Flow Meter Connector
Connect the sampler to Isco flow meters, Isco
flow loggers or 1640 Liquid Level Actuators by
attaching their connect cables to the flow meter
connector. Remember these guidelines:
• Both the sampler and the instrument
must be running a program.
• If using a flow meter or flow logger for
flow pacing, the sampler and the
instrument must both run programs with
flow pacing settings.
• If using a flow meter or flow logger for
trigger pacing, the sampler must run a
program with flow pacing settings, and
the instrument must run a program with
trigger pacing settings.
• When the sampler runs an event paced
program, it disregards pacing pulses
from a flow meter or flow logger.
However, the sampler continues to
monitor for enable signals.